US$489.00 · In stock Delivery: <= 4 days. True-PDF full-copy in English will be manually translated and delivered via email. JJG 440-2008: Verification Regulation of Industry Frequency Single-Phase Phase Meter Status: Valid JJG 440: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJG 440-2008 | English | 489 |
Add to Cart
|
4 days [Need to translate]
|
Verification Regulation of Industry Frequency Single-Phase Phase Meter
| Valid |
JJG 440-2008
|
JJG 440-1986 | English | 639 |
Add to Cart
|
5 days [Need to translate]
|
Verification Regulation of 50Hz Single-phase Phasometer
| Obsolete |
JJG 440-1986
|
PDF similar to JJG 440-2008
Basic data Standard ID | JJG 440-2008 (JJG440-2008) | Description (Translated English) | Verification Regulation of Industry Frequency Single-Phase Phase Meter | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A55 | Classification of International Standard | 17.220 | Word Count Estimation | 21,256 | Date of Issue | 2008-12-22 | Date of Implementation | 2009-06-22 | Older Standard (superseded by this standard) | JJG 440-1986 | Quoted Standard | JJF 1001-1998; GB/T 7676.1-1998; GB/T 7676.5-1998; GB/T 7676.9-1998; JB/T 9287-1999 | Regulation (derived from) | ?AQSIQ Announcement 2008 No.143 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard applies to the output range 10ns ~ 10000s time interval generator initial verification, follow-up tests and in-service inspection. |
JJG 440-2008: Verification Regulation of Industry Frequency Single-Phase Phase Meter---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Verification Regulation of Industry Frequency Single-Phase Phase Meter
State verification procedures People's Republic of China
Frequency single-phase phase meters
Posted 2008-12-22
2009-06-22 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Single-phase power frequency phase meter testing procedures
FrequencySing
Replacing JJG 440-1986
The regulations by the State Administration of Quality Supervision, Inspection and Quarantine approved on December 22, 2008, and from
From June 22, 2009 into effect.
Focal point. the National Technical Committee on Electromagnetic measurement
The main drafting units. School of Guizhou Province Measurement and Testing
Participated in the drafting unit. Henan Institute of Metrology
The measurement procedure commissioned the National Technical Committee is responsible for interpreting electromagnetic
The main drafters of this procedure.
Zhu Rui Yan (Guizhou Province Measurement and Testing Institute)
Wang Yan lamp (Guizhou Province Measurement and Testing Institute)
Lihong (Guizhou Province Measurement and Testing Institute)
Drafters participate.
Miao Hongwei (Henan Institute of Metrology)
Zhao Jun (Henan Institute of Metrology)
table of Contents
1 Scope (1)
2. Referenced Documents (1)
3 Overview (1)
4 metering performance requirements (1)
4.1 accuracy class (1)
4.2 basic error (1)
Lift variation 4.3 (2)
4.4 Non-rated load impact (2)
4.5 Effect of changing the working position (2)
4.6 Frequency Effect (3)
5 General technical requirements (3)
5.1 Appearance (3)
5.2 polar side button (3)
5.3 Power frequency withstand voltage (3)
5.4 Insulation resistance (3)
6 control of measuring instruments (3)
6.1 Verification condition (3)
6.2 test equipment (4)
7 test items (4)
8 test methods (5)
8.1 appearance inspection (5)
8.2 Direct comparison method (5)
8.3 basic error, lifting rated load variation and non-impact test (5)
8.4 Working position change impact test (7)
8.5 Frequency influence (7)
8.6 polar side button (8)
8.7 frequency voltage withstand test (8)
8.8 Insulation resistance (9)
9 test results processing and verification cycle (9)
9.1 Processing Verification Results (9)
9.2 test cycle (10)
Appendix A phase-frequency single-phase meter calibration record format (11)
APPENDIX B-frequency single-phase phase-table test certificate page format (13)
Appendix C frequency single-phase meter testing phase results in the notice page format (15)
D data rounding method appendix (16)
Single-phase power frequency phase meter testing procedures
1 Scope
The order applies to new manufacture, use and repair after the rated frequency 50Hz single-phase simulation phase pointer
Table (includes a phase angle meter and power factor meter, hereinafter referred to as phase meters) initial verification, subsequent verification and use of
test.
This procedure does not apply to electronic frequency phase meter, automatic recording type phase meter, phase compensation phase meters, non-public
Frequency phase meter, three-phase and phase table with this protocol does not match other types of phase measurement instruments.
2 Citations
The regulations cited in the following documents.
JJF1001-1998 common measurement terms and definitions
GB/T 7676.1-1998 Direct acting indicating analogue electrical measuring instruments and their accessories Part 1. Definitions and pass
With requirements (idtIEC 60051-1. 1984, Amend3.1997)
GB/T 7676.5-1998 Direct acting indicating analogue electrical measuring instruments and their accessories - Part 5. phase meters,
Particular requirements for power factor meters and synchronization indicator (idtIEC 60051-5. 1984)
GB/T 7676.9-1998 Direct acting indicating analogue electrical measuring instruments and their accessories Part 9. Recommended test
Test method (idtIEC 60051-9. 1988, Amend1.1994, Amend2.1995)
JB/T 9287-1999 phase meters, power factor meters and synchronization indicator
3 Overview
Single-phase meter phase is the phase difference between the current and the voltage of the AC circuit for measuring the frequency of the same special instrument,
Single-phase phase phase angle table into tables and power factor meter depending on the measurement function. Phase angle meter for measuring the same frequency
AC current and voltage phase angle difference between the phase angle is indicated, showing the value in degrees (°); power factor meter
For measuring the phase angle between current and voltage cosine, indicated that the power factor, dimensionless.
4 metering performance requirements
4.1 accuracy class
Single-phase phase meter accuracy level is divided into seven levels 0.5,1.0,1.5,2.0,2.5,3.0,5.0 each
Basic error should not exceed the level specified in table (single-phase basic error phase error table is represented by reference) 1.
4.2 basic error
4.2.1 The basic error of the phase angle of the table
Table 1, the level of accuracy and basic error limit
Accuracy class 0.5 1.0 1.5 2.0 2.5 3.0 5.0
Basic error limit/(%) ± 0.50 ± 1.0 ± 1.5 ± 2.0 ± 2.5 ± 3.0 ± 5.0
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJG 440-2008_English be delivered?Answer: Upon your order, we will start to translate JJG 440-2008_English as soon as possible, and keep you informed of the progress. The lead time is typically 2 ~ 4 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJG 440-2008_English with my colleagues?Answer: Yes. The purchased PDF of JJG 440-2008_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version JJG 440-2008?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version JJG 440-2008 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|