Home Cart Quotation About-Us
www.ChineseStandard.net
SEARCH

JJF 1383-2012 English PDF

US$369.00 ยท In stock
Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email.
JJF 1383-2012: Calibration Specification for Portable Blood Glucose Meters
Status: Valid
Standard IDUSDBUY PDFLead-DaysStandard Title (Description)Status
JJF 1383-2012369 Add to Cart 3 days Calibration Specification for Portable Blood Glucose Meters Valid

Similar standards

JJF 1033   JJF 1050   JJF 1030   

Basic data

Standard ID: JJF 1383-2012 (JJF1383-2012)
Description (Translated English): Calibration Specification for Portable Blood Glucose Meters
Sector / Industry: Metrology & Measurement Industry Standard
Word Count Estimation: 16,125
Quoted Standard: JJF 1001-2011; JJF 1059-1999; GB 6682-2008; GB/T 19634-2005; WS/T 226-2002
Regulation (derived from): AQSIQ Announcement No. 212 of 2012
Issuing agency(ies): General Administration of Quality Supervision, Inspection and Quarantine
Summary: This standard provides a portable blood glucose analyzer metrological characteristics, calibration conditions and calibration methods. This specification is for in vitro monitoring of blood sugar levels portable blood glucose analyzer.

JJF 1383-2012: Calibration Specification for Portable Blood Glucose Meters

---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for Portable Blood Glucose Meters People's Republic of China National Metrology Technical Specifications Portable blood glucose analyzer calibration specification Issued on. 2012-12-21 2013-03-21 implementation The State Administration of Quality Supervision, Inspection and Quarantine released Portable blood glucose analyzer calibration specification Focal point. the National Technical Committee on Clinical Measurement Main drafting unit. Hangzhou quality technical supervision detection hospital China Institute of Metrology Guangzhou Institute of Measurement and Detection Participated in the drafting unit. Ningbo Institute of Measurement and Testing Shanghai, a real medical equipment Co., Ltd. This specification commissioned the National Technical Committee of Clinical Medicine is responsible for interpretation of measurement The main drafters of this specification. Jiang Xueping (Hangzhou Quality and Technical Supervision and Inspection Institute) Xu Bei (China Institute of Metrology) Shao Lei (Hangzhou Quality and Technical Supervision and Inspection Institute) Huliang Yong (Guangzhou Institute of Measurement and Detection) Drafters participate. Yushan Cheng (Ningbo Institute of Measurement and Testing) Zhu Renming (Shanghai, a real medical equipment Co., Ltd.)

table of Contents

1 Scope (1) 2. References (1) 3 Terms and Definitions (1) 3.1 carryover (1) 4 Overview (1) 5 Metrological characteristics (1) 5.1 indication error (1) 5.2 Measurement repeatability (2) 5.3 linearity error (2) 5.4 carryover (2) Calibration Condition 6 (2) 6.1 environmental conditions (2) 6.2 standards and other measuring equipment (2) 7 calibration items and calibration methods (2) 7.1 appearance and working properly check (2) 7.2 preparation before calibration (2) 7.3 Indication error (2) 7.4 Measurement repeatability (3) 7.5 Linearity error (3) 7.6 carryover (3) 8 calibration results expression (4) 9 Recalibration interval (4) Appendix A portable blood glucose analyzer calibration record format (5) Appendix B portable blood glucose analyzer calibration certificate (within the page) format (6) Appendix C portable blood glucose analyzer Indication Error of Measurement Uncertainty (7) Portable blood glucose analyzer calibration specification

1 Scope

This specification defines the metrological characteristics of a portable blood glucose analyzer calibration conditions and calibration methods. The specification for the body External monitoring of blood glucose content portable blood glucose analyzer. 2. References JJF 1001-2011 common measurement terms and definitions JJF 1059-1999 Measurement Uncertainty and representation GB 6682-2008 analytical laboratory use specifications and test methods GB/T 19634-2005 vitro diagnostic test systems for self blood glucose monitoring system General technical requirements WS/T 226-2002 portable blood glucose meter blood glucose measurement guide For dated references, only the dated edition applies to this specification; undated reference documents Member, the latest edition (including any amendments) applies to this specification.

3 Terms and Definitions

3.1 carryover carry-overrate By the measurement system will carry a test sample reaction to detect the analyte another sample of the reaction discrete amount by This error affected the performance of another quantity of the test sample. Usually taken given the known nominal value of high and low sample Measured value of the sample cross many times, calculated using a specific formula.

4 Overview

Portable blood glucose analyzer according to the measure of sugar technology can be divided electrochemical test and the light reflective technology test two categories. The former is an enzyme reacts with glucose to produce an electronic current through the counting device, read the number of electrons; microcontroller according to Correspondence between the current size and the blood glucose concentration in the blood glucose concentration is calculated and displayed. The latter is by Contact with the blood test reagent block blood glucose test strips, and color change agent blocks react portable blood glucose By irradiating the test wells analyzer reagent strip block strip on the platform, and measuring the reflected light, thereby obtaining a reagent strip Built-in microcontroller based on the color of the corresponding block and reagent strip blood glucose concentration; changes in color blocks Relationship between blood glucose concentration values calculated and displayed. Portable blood glucose analyzer by the analysis system, microcontrollers, data storage, LCD display, data output interface Ports and other components. Rapid measurement portable blood glucose analyzer is mainly used in fresh capillary whole blood glucose content can also be used Venous whole blood, rapid measurement of serum (plasma) of the glucose content.

5 Metrological characteristics

5.1 Indication error
......
Image     

Tips & Frequently Asked Questions:

Question 1: How long will the true-PDF of JJF 1383-2012_English be delivered?

Answer: Upon your order, we will start to translate JJF 1383-2012_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time.

Question 2: Can I share the purchased PDF of JJF 1383-2012_English with my colleagues?

Answer: Yes. The purchased PDF of JJF 1383-2012_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet.

Question 3: Does the price include tax/VAT?

Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countries

Question 4: Do you accept my currency other than USD?

Answer: Yes. If you need your currency to be printed on the invoice, please write an email to Sales@ChineseStandard.net. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.