Standard ID | JJF 1310-2011 (JJF1310-2011) | Description (Translated English) | Calibration Specification for Electronic Plug Gauges | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A52 | Classification of International Standard | 17.040 | Word Count Estimation | 17,112 | Date of Issue | 2011-09-14 | Date of Implementation | 2011-12-14 | Quoted Standard | JJG 894-1995; JJF 1001-1998; JJF 1059-1999; JB/T 7429-2006 | Issuing agency(ies) | State Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard applies to measuring range of 130mm diameter 6mm-diameter two-point calibration of electronic plug, this plug outside the scope of electronics and electronic plug-point calibration can be calibrated with reference to the specification. |
JJF 1310-2011
Calibration Specification for Electronic Plug Gauges
People's Republic of China National Metrology Technical Specifications
Electronic plug gauge calibration specification
Issued on. 2011-09-14
2011-12-14 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Electronic plug gauge calibration specification
This specification by the State Administration of Quality Supervision, Inspection and Quarantine approved on September 14, 2011, and from
From December 14, 2011 implementation.
NFP. National Engineering Geometrical Parameters Measurement Technology Committee
Main drafting unit. Guangxi Institute of measurement testing
Heilongjiang Provincial Academy of verification test
Participated in the drafting unit. Meter Zhongyuan Ltd.
Sanmenxia in measuring instrument Co., Ltd.
Jiangsu Province Institute of Metrology
This specification project commissioned by the National geometric parameter measurement technology committee responsible for the interpretation
The main drafters of this specification.
Full Yi-chi (Guangxi Institute of measurement detection)
Lu Rui (Guangxi Institute of measurement detection)
Liang Yuhong (Heilongjiang Provincial Academy of verification test)
Drafters participate.
Zhang Hongfei (Zhongyuan Measuring Instruments Co., Ltd.)
Ji Yihong (Sanmenxia in measuring instrument Co., Ltd.)
Goal Orientation (Jiangsu Province Institute of Metrology)
table of Contents
1 Scope (1)
2. Referenced Documents (1)
3 Terms and Definitions (1)
4 Overview (1)
5 Metrological characteristics (2)
5.1 Calibration with a standard ring gauge diameter and the amount of variation in diameter (2)
5.2 Work space (2)
Measuring force 5.3 (2)
5.4 probing surface and guide face surface roughness (2)
5.5 Zero balance (3)
5.6 Zero range (3)
5.7 Probe alignment error (3)
5.8 Probe symmetry error (3)
5.9 Indication error (3)
5.10 Indication variability (3)
Drift 5.11 (3)
6 calibration conditions (4)
Environmental conditions 6.1 (4)
6.2 measurement standards (4)
7 calibration items and calibration methods (4)
7.1 calibration items (4)
7.2 Calibration method (5)
8 calibration results expression (6)
9 Recalibration interval (6)
Appendix A electronic plug gauge indication error measurement Uncertainty Evaluation (7)
Appendix B calibration certificate content and in-page format (10)
Electronic plug gauge calibration specification
1 Scope
This standard applies to the measurement range of φ6mm ~ two electronic plug gauge φ130mm calibration, this range
Electronic plug gauge plug gauge and three-point electronic calibration can be outside of this specification with reference to the calibration.
2 Citations
This specification references the following documents.
JJG894-1995 standard ring gauge
JJF 1001-1998 common measurement terms and definitions
JJF 1059-1999 Evaluation and Expression of Uncertainty in Measurement
JB/T 7429-2006 electronic plug gauge
When using this specification it should be noted that using the currently valid version of the documents cited.
3 Terms and Definitions
3.1 Work space workingclearance
Deviation of the difference between the size of the maximum diameter of the hole diameter and the measured electronic plug gauge guide sleeve.
3.2 Probe alignment errors contactvariationagainstcerterline
Two plug connections departing from the probe contacts the measurement error plug wire guide sleeve shaft diameter generated.
3.3 Probe symmetric error contactsymmetricdifferent
Plug two probes lever hinge mechanism with respect to the central axis of the asymmetric shift plug arising from measurement error.
3.4 Zero balance nulbalance
The pointing device in the conversion range gear, showing the value of each block of the zero offset.
3.5 Indication variability instability
In the case of the measurement conditions do not make any changes, it is measured against the same number of measurements between the maximum result
difference.
4 Overview
Electronics is a plug-contact measurement aperture comparative measurement instruments, will be measured by the inductive sensor aperture size changes
Into electrical signals, and then by indicating means for indicating the measured aperture size deviation. Electronic plug by the plug body, indicating means
(Pointer table, DRO or electron column) and the composition of the calibration standard ring gauge. Electronic plug according to the number of probes can be divided into two
Point and three-point, two-point common electronic plug gauge its shape structure shown in Figure 1.
......
|