Delivery: <= 6 days. True-PDF full-copy in English will be manually translated and delivered via email. GB 8599-2023: Technical requirements for large steam sterilizers Status: Valid GB 8599: Historical versions
Basic dataStandard ID: GB 8599-2023 (GB8599-2023)Description (Translated English): Technical requirements for large steam sterilizers Sector / Industry: National Standard Classification of Chinese Standard: C47 Classification of International Standard: 11.080.10 Word Count Estimation: 38,340 Date of Issue: 2023-09-08 Date of Implementation: 2026-10-01 Older Standard (superseded by this standard): GB 8599-2008 Issuing agency(ies): State Administration for Market Regulation, China National Standardization Administration GB 8599-2023: Technical requirements for large steam sterilizers---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.ICS 11.080.10 CCSC47 National Standards of People's Republic of China Replace GB 8599-2008 Technical requirements for large pressure steam sterilizers Published on 2023-09-08 Implemented on 2026-10-01 State Administration for Market Regulation Released by the National Standardization Administration Committee Table of contentsPreface III 1 Scope 1 2 Normative reference documents 1 3 Terms and Definitions 1 4 Category 3 5 Requirements 3 6 Test methods 10 Appendix A (Informative) Environmental Impact 19 Appendix B (informative) Qualification standards for sterilizers of the same series20 Appendix C (informative) Allowable range of temperature and time during small load temperature test 21 Appendix D (Normative) Test instruments, equipment and materials 22 Appendix E (informative) Information and documents provided by the manufacturer28 Appendix F (informative) Normal working conditions 30 Reference 33ForewordThis document complies with the provisions of GB/T 1.1-2020 "Standardization Work Guidelines Part 1.Structure and Drafting Rules of Standardization Documents" Drafting. This document replaces GB 8599-2008 "Technical Requirements for Large Steam Sterilizers Automatic Control Type". Compared with GB 8599-2008, except In addition to structural adjustments and editorial changes, the main technical changes are as follows. a) Changed the scope of the standard (see Chapter 1, Chapter 1 of the.2008 edition); b) Changes in terms and definitions (see Chapter 3, Chapter 3 of the.2008 edition); c) Basic parameters are deleted (see 4.2 of the.2008 version); d) Changed the requirements for double-door sterilizers (see 5.1.1.2, 5.4.2 of the.2008 edition); e) Changed the requirements for testing connectors (see 5.1.1.3, 5.4.3 of the.2008 version); f) Normal working conditions are deleted (see 5.1 of the.2008 version); g) Changed the requirements for structure and sterilization chamber size (see 5.2.1.1, 5.2.3 of the.2008 edition); h) The requirements for appearance are deleted (see 5.2.1 and 5.2.2 of the.2008 version); i) Removed material requirements (see 5.3 of the.2008 version); j) The requirement for interlocking devices is deleted (see 5.4.1.2 of the.2008 edition); k) The requirements for thermal insulation materials are deleted (see 5.4.4 of the.2008 edition); l) Changed the requirements for pipelines (see 5.2.2.1, 5.5.1 of the.2008 version); m) Changed the requirements for steam generating devices (see 5.2.2.2, 5.5.2 of the.2008 edition); n) Changed the requirements for air filters (see 5.2.2.3, 5.5.3 of the.2008 edition); o) Changes to the content of measuring systems, display and recording devices (see 5.2.3, 5.6 of the.2008 version); p) Changed the requirements for control systems (see 5.2.4, 5.7 of the.2008 edition); q) Changed the criteria for BD test (see 5.3.1.1, 5.8.1.1 of the.2008 version); r) The relevant requirements, test methods and appendices for the killing of biological indicators have been changed (see 5.3.2, 6.4.2, D.3,.2008 edition 5.8.2, 6.8.2, F.8); s) Added the calculation method of the theoretical temperature of saturated steam [see 5.3.3.1c)]; t) Changed the requirements for small load temperature [see 5.3.3.1d),.2008 version of 5.8.3.2d)]; u) Changed the requirements for full load temperature [see 5.3.3.2d),.2008 version of 5.8.3.3d)]; v) Delete the relevant requirements for marking (see 5.12 of the.2008 version); w) Changed the requirements for cavity load test (see 6.4.1.2, 6.8.1.2 of the.2008 edition); x) Added relevant requirements for wireless temperature sensors in temperature performance testing (see 6.4.3, 6.8.3 of the.2008 version); y) Changed the test requirements for noise test (see 6.4.6, 6.9 of the.2008 edition); z) Change the normative appendix "Temperature and time allowable range during small load temperature test" to an informative appendix (see Appendix C, Appendix E of the.2008 edition); aa) The requirements of the normative appendix "Testing instruments, equipment and materials" have been changed (see Appendix D, Appendix F of the.2008 version). Please note that some content in this document may be subject to patents. The publisher of this document assumes no responsibility for identifying patents. This document is proposed and managed by the National Medical Products Administration. The previous versions of this document and the documents it replaces are as follows. ---First published in 1988 as GB 8599-1988 "Technical Conditions for Automatic Control of Pressure Steam Sterilizers"; ---In the first revision in.2008, the contents of GB 8600-1988 "Sterilization Effect Test Method of Pressure Steam Sterilizer" were integrated. content, the name was changed to "Technical Requirements for Large Steam Sterilizer Automatic Control Type"; ---This is the second revision. Technical requirements for large pressure steam sterilizers1 ScopeThis document defines the terms and definitions of large pressure steam sterilizers (hereinafter referred to as "sterilizers"), stipulates classifications, requirements, and describes experiment method. This document is applicable to sterilization chambers with side openings that can load at least one sterilization unit or a volume of not less than 60L and a vacuum function. Bacteria. This sterilizer is mainly used for sterilizing medical and health care products and their accessories (excluding liquid items, flammable and explosive items, biological waste, human sterilization of body tissue). Note. Sterilizers designed and produced according to this document also need to consider the impact of environmental factors in the product life cycle. See Appendix A for environmental impact factors.2 Normative reference documentsThe contents of the following documents constitute essential provisions of this document through normative references in the text. Among them, the dated quotations For undated referenced documents, only the version corresponding to that date applies to this document; for undated referenced documents, the latest version (including all amendments) applies to this document. GB/T 3768 Acoustic sound pressure method determines the sound power level and sound energy level of noise sources using a simplified method of enveloping the measuring surface above the reflecting surface. Yifa GB/T 3785 (all parts) Electroacoustic sound level meter GB 4793.1 Safety requirements for electrical equipment for measurement, control and laboratory use Part 1.General requirements GB 4793.4 Safety requirements for electrical equipment for measurement, control and laboratory use Part 4.Sterilizers for processing medical materials and special requirements for washer-disinfectors GB/T 16839.1 Thermocouples Part 1.Electromotive force specifications and tolerances GB/T 18268.1 Electromagnetic compatibility requirements for electrical equipment for measurement, control and laboratory use Part 1.General requirements GB 18278.1-2015 Moist heat sterilization of health care products Part 1.Development, validation and routine of sterilization process of medical devices Control requirements GB 18281.3 Biological indicators for sterilization of medical and health care products Part 3.Biological indicators for moist heat sterilization GB 18282.1 Chemical indicators for sterilization of health care products Part 1.General principles GB 18282.3 Chemical Indicators for Sterilization of Healthcare Products Part 3.Class II Indicators for BD Class Vapor Penetration Testing physical system GB/T 19633 (all parts) Packaging of terminally sterilized medical devices GB/T 19971-2015 Terminology for sterilization of medical and healthcare products GB/T 30121 Industrial platinum thermal resistor and platinum temperature sensing element3 Terms and definitionsThe following terms and definitions as defined in GB 18278.1-2015 and GB/T 19971-2015 apply to this document. 3.1 Absolute pressureabsolutepressure The pressure with absolute vacuum as the zero reference. ......Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of GB 8599-2023_English be delivered?Answer: Upon your order, we will start to translate GB 8599-2023_English as soon as possible, and keep you informed of the progress. The lead time is typically 4 ~ 6 working days. The lengthier the document the longer the lead time.Question 2: Can I share the purchased PDF of GB 8599-2023_English with my colleagues?Answer: Yes. The purchased PDF of GB 8599-2023_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet.Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to Sales@ChineseStandard.net. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.Question 5: Should I purchase the latest version GB 8599-2023?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version GB 8599-2023 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically. |