HOME   Cart(0)   Quotation   About-Us Tax PDFs Standard-List Powered by Google www.ChineseStandard.net Database: 189759 (6 Oct 2024)

YY/T 1533-2017 related PDF English

YY/T 1533-2017 (YY/T1533-2017, YYT 1533-2017, YYT1533-2017) & related versions
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)See DetailStatusSimilar PDF
YY/T 1533-2017English150 Add to Cart 0-9 seconds. Auto delivery. Automatic time-resolved fluorescence counter YY/T 1533-2017 Valid YYT 1533-2017
Buy with any currencies (Euro, JPY, KRW...): YY/T 1533-2017    Preview this PDF: YY/T 1533-2017



YY/T 1533-2017: PDF in English (YYT 1533-2017)
YY/T 1533-2017 YY PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE’S REPUBLIC OF CHINA ICS 11.100 C 44 Automatic time-resolved fluorescence counter ISSUED ON: MARCH 28, 2017 IMPLEMENTED ON: APRIL 01, 2018 Issued by: China Food and Drug Administration Table of Contents Foreword ... 3  1 Scope ... 4  2 Normative references ... 4  3 Requirements ... 5  4 Test method ... 6  5 Markings, labels, user instructions ... 11  6 Packaging, transportation, storage ... 11  References ... 12  Automatic time-resolved fluorescence counter 1 Scope This standard specifies the requirements, test methods, markings, labels, and instructions for use, packaging, transportation, storage of automatic time- resolved fluorescence counter (hereinafter referred to as analyzers). This standard applies to automatic time-resolved fluorescence counter. 2 Normative references The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard. GB/T 191 Packaging - Pictorial marking for handling of goods GB 4793.1 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements GB 4793.9 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 9: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes GB/T 14710 Environmental requirement and test methods for medical electrical equipment GB/T 18268.1 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements GB/T 18268.26 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 26: Particular requirements - In vitro diagnostic (IVD) medical equipment GB/T 29791.3 In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 3: In vitro diagnostic instruments for professional use YY 0648 Safety requirements for electrical equipment for measurement control and laboratory use - Part 2-101: Particular requirements for in vitro If pipetting volume of ≥ 10 µL and < 50 µL, the relative deviation of loading accuracy is ≤ ±5%; if pipetting volume of ≥ 50 µL, the relative deviation of loading accuracy is ≤ ±3%. 3.3.2 Precision of sample loading If pipetting volume of ≥ 10 µL and < 50 µL, the precision of sample loading CV is ≤ 5%. If pipetting volume of ≥ 50 µL, the precision of sample loading CV is < 2%. 3.3.3 Contamination by sample loading needle Carrying contamination rate is < 0.01%. 3.3.4 Residual amount of lotion The residual amount of lotion is < 3 μL/hole. 3.4 Temperature accuracy and fluctuation The temperature accuracy of the incubation chamber is within ±1.5 °C. The temperature fluctuation is within ±1.5 °C. 3.5 Repeatability of clinical test The in-batch repeatability CV of the kit is ≤ 15%. 3.6 Environmental test requirements It shall meet the requirements of the applicable clauses in GB/T 14710. 3.7 Electrical safety requirements It shall meet the requirements of the applicable clauses in GB 4793.1, GB 4793.9, YY 0648. 3.8 Electromagnetic compatibility requirements It shall meet the requirements of the applicable clauses of GB/T 18268.1 and GB/T 18268.26. 4 Test method 4.1 Appearance Visual inspection with normal vision or corrected vision under natural light shall meet the requirements of 3.1. error of 10-10 mol/L europium standard solution. 4.2.5 Stability After the analyzer is turned on and in a stable working state, take one set of 200 μL 10-10 mol/L europium standard solution to repeat the test 10 times. Calculate the average value of the measurement results. Then re-start the machine after 4 h and 8 h to repeat the test 10 times; calculate the average value of the measurement result. Take the first measurement result as the reference value; calculate the relative offset (α, %) according to formula (2), which shall meet the requirements of 3.2.5. Wherein: - The average of the measured values in the 4th and 8th hours; - The average value of the initial measured value. 4.3 Sample loading and cleaning system 4.3.1 Sample loading accuracy The following methods can be used for verification: a) Take 12 clean microwell cups; use a calibrated analytical balance (accuracy is ±0.1 mg) to weigh each empty cup; record it as W0i (where i = 1, …, 12); b) Put the weighed empty cups into the first row of flat-bottomed microplates in turn; add 100 μL of 0.9% saline to the first row of micro-bottomed cups of flat-bottomed microplates; weigh each microwell cup and record it as W1i (where i = 1,…, 12); c) Calculate the result of (W1i - W0i) and convert it to volume, which is the sample loading volume per hole; d) Calculate the relative deviation between the actual sample loading volume per hole and the set sample loading volume, which is the sample addition accuracy and shall meet the requirements of 3.3.1. Note: It can also be verified according to the test method provided by the manufacturer. formula (3) and shall meet the requirements of 3.3.3. Wherein: c - Concentration value of high concentration sample. 4.3.4 Residual volume of lotion The following methods can be used for verification: a) Take a clean microplate and use a calibrated ten-thousandth balance to weight it; record it as M1; b) Put the weighed microplate on any path of plate washing platform; run the plate washing twice; soak it for 2 s; c) Weigh the microplate after washing the plate and record it as M2; d) Calculate the average value of the residual volume of each hole; convert it into volume, which is the residual volume of the washing solution and shall meet the requirements of 3.3.4. 4.4 Temperature accuracy and fluctuation The following methods can be used for verification: a) Put the temperature sensor probe of the thermometer (accuracy 0.1 °C) into any incubation chamber; b) Set the temperature of the incubation chamber to be tested to 37 °C and run it continuously for 1 h with continuous shaking; c) Record the temperature from the 16th minute; record once every 1 minute; record 10 times in a row; d) Calculate the average value as well as the difference between the maximum and minimum temperatures in the chamber to be tested. The difference between the average value of the chamber temperature and the set temperature value is the temperature accuracy; the half of the difference between the maximum value and the minimum value of the chamber temperature is the temperature fluctuation. The result shall meet the requirements of 3.4. 4.5 Repeatability of clinical test ......

BASIC DATA
Standard ID YY/T 1533-2017 (YY/T1533-2017)
Description (Translated English) Automatic time-resolved fluorescence counter
Sector / Industry Medical Device & Pharmaceutical Industry Standard (Recommended)
Classification of Chinese Standard C44
Classification of International Standard 11.100
Word Count Estimation 8,860
Date of Issue 2017-03-28
Date of Implementation 2018-04-01
Drafting Organization Beijing Medical Device Inspection Institute, Suzhou New Wave Biological Technology Co., Ltd., Shenzhen Ai Kang Biological Technology Co., Ltd.
Administrative Organization National Medical Clinical Laboratory and in vitro diagnostic system standardization Technical Committee (SAC/TC 136)
Regulation (derived from) China Food & Drug Administration Announcement 2017 No. 38
Proposing organization China Food and Drug Administration
Issuing agency(ies) State Food and Drug Administration