YY/T 1510-2017 (YY/T1510-2017, YYT 1510-2017, YYT1510-2017) & related versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | See Detail | Status | Similar PDF |
YY/T 1510-2017 | English | 90 |
Add to Cart
|
0-9 seconds. Auto delivery.
|
Medical plasma virus inactivated device
|
YY/T 1510-2017
| Valid |
YYT 1510-2017
|
Buy with any currencies (Euro, JPY, KRW...): YY/T 1510-2017 Preview this PDF: YY/T 1510-2017
YY/T 1510-2017
YY
PHARMACEUTICAL INDUSTRY STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
ICS 11.040.20
C 31
Medical Plasma Virus Inactivated Device
ISSUED ON. MARCH 28, 2017
IMPLEMENTED ON. APRIL 1, 2018
Issued by. China Food and Drug Administration
Table of Contents
Foreword ... 3
Introduction ... 4
1 Scope ... 5
2 Normative References ... 5
3 Terms and Definitions ... 5
4 Requirements... 6
5 Test Methods ... 7
6 Inspection Rules ... 8
7 Marking, Instruction for Use ... 9
8 Packaging ... 10
Introduction
This Standard provides basic technical guidance to medical plasma virus inactivated
device of different light sources, design structures and specification models.
Medical plasma virus inactivated device is a device of preparing clinically infused virus
inactivated plasma for blood collection and supply institutions, and clinical medical
institutions. It is used cooperatively with methylene blue virus inactivated device.
Medical plasma virus inactivated device is mainly applicable to photochemical method
for plasma virus inactivation technology. In other words, after adding an appropriate
amount of virus inactivating agent---methylene blue to plasma and placing it in the
device, plasma can receive certain amount of illumination at a certain temperature.
Under the illumination, sway the plasma, so that the plasma that contains methylene
blue can receive uniform and effective illumination energy, which will reach the effect
of virus inactivation. Medical plasma virus inactivated device is an exclusive supporting
device that guarantees safe, effective and reliable plasma virus inactivation through
the photochemical method.
Medical plasma virus inactivated device provides appropriate illumination conditions to
plasma virus inactivation through the photochemical method. Effective illumination
conditions for virus inactivation depend on the variety of light source and the intensity
of illumination that is emitted from a specific waveband, and illumination energy that
plasma receives. There are two commonly seen light sources of virus inactivated
device in the market. florescence and LED, and the duration of illumination also varies
with the intensity of illumination. In consideration of these, this Standard specifies that
illumination parameters stipulated by manufacturers shall be satisfied in tests with
methods provided by the manufacturers. However, satisfying these parameters does
not signify that it can reach an ideal effect of virus inactivation. Manufacturers shall
take the responsibility of verifying the validity of these parameters and confirming the
effect of virus inactivation in accordance with relevant national-level stipulations.
Furthermore, manufacturers of medical plasma virus inactivated device shall confirm
that there is no obvious adverse influence on blood component quality and function
and verify infusion safety after virus inactivation. Thus, after virus inactivation, blood
component can comply with relevant national-level stipulations.
Medical Plasma Virus Inactivated Device
1 Scope
This Standard specifies the general requirements, test methods, inspection rules,
marking, instruction for use and packaging of medical plasma virus inactivated device
(hereinafter referred to as virus inactivated device).
This Standard is applicable to medical plasma virus inactivated device that is matched
with virus inactivated device that is stipulated in YY 0765.1.
2 Normative References
The following documents are indispensable to the application of this Standard. In terms
of references with a specified date, only versions with a specified date are applicable
to this Standard. The latest version (including all the modifications) of references
without a specified date is also applicable to this Standard.
GB 4793.1 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use - Part 1. General Requirements
GB/T 14710 Environmental Requirement and Test Methods for Medical Electrical
Equipment
GB/T 18268.1 Electrical Equipment for Measurement, Control and Laboratory Use
- EMC Requirements - Part 1. General Requirements
YY/T 0466.1 Medical Devices - Symbols to be Used with Medical Device Labels,
Labelling and Information to be Supplied - Part 1. General Requirements
YY 0765.1 Sets for Inactivation of Viruses in Blood and Blood Components for
Single Use. Part 1. Sets for Virus Photodynamic Inactivation with Methylene Blue
3 Terms and Definitions
The following terms and definitions are applicable to this Standard.
3.1 Stable Operating State
Stable operating state means after virus inactivated device is activated, indicated
temperature reaches 2 °C ~ 8 °C; illumination intensity reaches the range explicitly
instructed by manufacturers.
3.2 Effective Inactivation Area
5.3 Illumination Intensity
Conduct the test in accordance with methods stipulated by manufacturers.
5.4 Smoothness and Reliability of Load Shelf
Under the maximum load, conduct visual inspection of whether the load shelf in the
virus inactivated device can operate smoothly and reliably; whether the extraction is
flexible; whether there are obvious sway and distortion.
5.5 Alarm Temperature
When the practical simulated temperature and illumination intensity exceed the
limitation, use sound level meter to measure alarm sound level at a distance of 1 m
from the front panel and 1 m from the ground.
5.6 Time Setting
Use a timer to measure the practical operating hours of virus inactivated device;
calculate the deviation from the previously set time.
6 Inspection Rules
6.1 Inspection Type
Inspection of virus inactivated device shall be divided into exit-factory inspection and
model inspection.
6.2 Exit-factory Inspection
Exit-factory inspection items of each virus inactivated device (including process
inspection and/or ultimate inspection) shall at least include 4.2 ~ 4.9.
6.3 Model Inspection
6.3.1 Model inspection shall be conducted under any of the following circumstances.
a) When new products are put into production;
b) When there are significant changes in structure, critical components and
technology;
c) When production is resumed after production suspension and rectification;
d) When there are significant differences between the result of exit-factory
inspection result and the last model inspection;
e) When it is stipulated by contract or requested by management department.
......
Standard ID | YY/T 1510-2017 (YY/T1510-2017) | Description (Translated English) | Medical plasma virus inactivated device | Sector / Industry | Medical Device & Pharmaceutical Industry Standard (Recommended) | Classification of Chinese Standard | C31 | Classification of International Standard | 11.040.20 | Word Count Estimation | 6,627 | Date of Issue | 2017-03-28 | Date of Implementation | 2018-04-01 | Drafting Organization | Shanghai Blood Center, Shandong Province, medical device product quality inspection center | Administrative Organization | National Medical Infusion Equipment Standardization Technical Committee (SAC/TC 106) | Proposing organization | National Medical Infusion Equipment Standardization Technical Committee (SAC/TC 106) | Issuing agency(ies) | State Food and Drug Administration | Standard ID | () | Description (Translated English) | (Technical specifications for ceramic capacitive sensor type partial discharge monitoring devices of 35kV and below) | Sector / Industry | Chinese Industry Standard | Date of Issue | 2023-10-11 | Date of Implementation | 2024-04-11 | Issuing agency(ies) | National Energy Board | Standard ID | () | Description (Translated English) | (Technical specifications for ceramic capacitive sensor type partial discharge monitoring devices of 35kV and below) | Sector / Industry | Chinese Industry Standard | Date of Issue | 2023-10-11 | Date of Implementation | 2024-04-11 | Issuing agency(ies) | National Energy Board |
|