YY/T 0806-2010 (YY/T0806-2010, YYT 0806-2010, YYT0806-2010) & related versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | See Detail | Status | Similar PDF |
YY/T 0806-2010 | English | 190 |
Add to Cart
|
0-9 seconds. Auto delivery.
|
Polycarbonate material for manufacture of infusion, transfusion and injection equipments for medical use and other medical devices
|
YY/T 0806-2010
| Valid |
YYT 0806-2010
|
Buy with any currencies (Euro, JPY, KRW...): YY/T 0806-2010 Preview this PDF: YY/T 0806-2010
YY/T 0806-2010
PHARMACEUTICAL INDUSTRY STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
ICS 11.040.20
C 31
Polycarbonate material for manufacture of infusion,
transfusion and injection equipments for medical use
and other medical devices
ISSUED ON: DECEMBER 27, 2010
IMPLEMENTED ON: JUNE 01, 2012
Issued by: China Food and Drug Administration
Table of Contents
Foreword ... 3
Introduction ... 4
1 Scope ... 5
2 Normative references ... 5
3 Requirements ... 6
4 Test method ... 7
5 Marking ... 9
6 Packaging and storage ... 9
References ... 10
Polycarbonate material for manufacture of infusion,
transfusion and injection equipments for medical use
and other medical devices
1 Scope
This Standard specifies the requirements for polycarbonate material for
manufacture of infusion, transfusion and injection equipments for medical use
and other medical devices
This Standard does not apply to polycarbonate copolymers.
2 Normative references
The following referenced documents are indispensable for the application of
this document. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any
amendments) applies.
GB/T 1033.1-2008, Plastics - Methods for determining the density of non-
cellular plastics - Part 1: Immersion method liquid pyknometer method and
titration method (ISO 1183-1:2004, IDT)
GB/T 1843-2008, Plastics - Determination of izod impact strength
GB/T 2547-2008, Plastic resins - Sampling
GB/T 3682, Determination of the melt mass-flow rate (MFR) and the melt
volume-flow rate (MVR) of thermoplastics
GB/T 9352-2008, Plastic - Compression moulding of test specimens of
thermoplastic materials
GB/T 14233.1-2008, Test methods for infusion transfusion injection
equipments for medical use - Part 1: Chemical analysis methods
GB/T 16886.1, Biological evaluation of medical devices - Part 1: Evaluation
and testing within a risk management process
GB/T 2918-1998, Plastics - Standard atmospheres for conditioning and
testing
Visual inspection.
4.4 Physical properties
4.4.1 Melt volume-flow rate
Take particle samples according to GB/T 3682.
4.4.2 Density
Take particle samples according to method B in GB/T 1033.1-2008.
4.4.3 Izod notched impact strength
Take samples prepared in 5.1 and carry out according to method A in GB/T
1843-2008.
4.4.4 Radiation-resistant materials
For radiation-resistant sterilization polycarbonate materials for medical use,
take a 3 mm-thick test piece; after 30 kGy radiation, place it under natural light
conditions for 15 days; test the yellow index before and after radiation according
to HG/T 3862.
4.5 Chemical properties of dissolved substances
4.5.1 Preparation of test solution
Weigh particle samples and use the class-2 test water as specified in GB/T
6682-2008 to rinse them; add water in a ratio of 1:5 of mass (g) to water (mL);
seal with a lid; place them in a pressure steam sterilizer, at 121 °C ± 1 °C of
saturated steam, for 30 min; separate the sample from the solution; cool to room
temperature. Take the same batch of water to prepare blank control solution by
the same method.
4.5.2 Reducing substance
Take the test solution and control solution, which are prepared according to
4.5.1, and carry out according to 5.2.2 of GB/T 14233.1-2008.
4.5.3 pH value
Take the test solution and control solution, which are prepared according to
4.5.1, and carry out according to 5.4.1 of GB/T 14233.1-2008.
4.5.4 Evaporation residue
Take the test solution and control solution, which are prepared according to
4.5.1, and carry out according to 5.5 of GB/T 14233.1-2008.
4.5.5 Metal ion
Take the test solution that is prepared according to 4.5.1, and carry out
according to 5.6.1 of GB/T 14233.1-2008 and/or the atomic absorption
spectrophotometry (AAS) or equivalent method.
4.5.6 UV absorbance
Take the test solution that is prepared according to 4.5.1, and carry out in the
wavelength range of 250 nm ~ 320 nm according to the provisions of 5.7 of
GB/T 14233.1-2008.
5 Marking
The marking on the outer packaging bag of the medical polycarbonate material
product shall include the following contents:
a) Product name;
b) Manufacturer's name or trademark;
c) Model;
d) Batch number;
e) Net weight.
6 Packaging and storage
6.1 Packaging
Polycarbonate materials for medical use shall be at least double-packed, and
the packaging shall ensure that the product will not be contaminated during
transportation and storage.
6.2 Storage
Polycarbonate materials for medical use shall be stored in a ventilated, dry and
clean warehouse with well-maintained firefighting facilities. Store away from
heat sources and avoid direct sunlight.
......
Standard ID | YY/T 0806-2010 (YY/T0806-2010) | Description (Translated English) | Polycarbonate material for manufacture of infusion, transfusion and injection equipments for medical use and other medical devices | Sector / Industry | Medical Device & Pharmaceutical Industry Standard (Recommended) | Classification of Chinese Standard | C31 | Classification of International Standard | 11.040.20 | Word Count Estimation | 10,16 | Date of Issue | 2010-12-27 | Date of Implementation | 2012-06-01 | Quoted Standard | GB/T 1033.1-2008; GB/T 1843-2008; GB/T 2547-2008; GB/T 3682; GB/T 9352-2008; GB/T 14233.1-2008; GB/T 16886.1; GB/T 2918-1998; GB/T 6682; HG/T 3862 | Adopted Standard | ASTM F997-1998a (2003), NEQ | Drafting Organization | Shandong Province Quality Inspection Center of Medical Devices | Administrative Organization | National Technical Committee of Standardization for medical infusion | Regulation (derived from) | State Food and Drug Administration Notice 2010 No. 97 | Proposing organization | National Standardization Technical Committee appliances Infusion | Issuing agency(ies) | China Food and Drug Administration | Summary | This standard specifies the medical infusion, blood transfusion, injection and other medical equipment polycarbonate Compound requirements. This standard does not apply to a polycarbonate copolymer. |
|