HOME   Cart(0)   Quotation   About-Us Tax PDFs Standard-List Powered by Google www.ChineseStandard.net Database: 189760 (12 Oct 2024)

YY/T 0654-2017 related PDF English (YYT0654-2008)

YY/T 0654-2017 (YY/T0654-2017, YYT 0654-2017, YYT0654-2017) & related versions
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)See DetailStatusSimilar PDF
YY/T 0654-2017English145 Add to Cart 0-9 seconds. Auto delivery. Automatic chemistry analyzer YY/T 0654-2017 Valid YYT 0654-2017
YY/T 0654-2008English519 Add to Cart 3 days Automatic Chemistry Analyzer YY/T 0654-2008 Obsolete YYT 0654-2008
Buy with any currencies (Euro, JPY, KRW...): YY/T 0654-2017    Preview this PDF: YY/T 0654-2017



YY/T 0654-2017 YY PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE’S REPUBLIC OF CHINA ICS 11.100 C 44 Replacing YY/T 0654-2008 Automatic chemistry analyzer ISSUED ON. MARCH 28, 2017 IMPLEMENTED ON. APRIL 1, 2018 Issued by. China Food and Drug Administration of China Table of Contents Foreword ... 3  1 Scope ... 5  2 Normative references ... 5  3 Terms and definitions ... 6  4 Classification ... 6  5 Requirements ... 7  6 Test methods ... 9  7 Marks and instructions for use ... 15  8 Packaging, transport and storage ... 15  Annex A (normative) Preparation method of 50g/L sodium nitrite solution ... 17  Annex B (informative) Density of pure water at different temperatures at standard atmospheric pressure ... 18  Foreword This Standard was drafted in accordance with the rules given in GB/T 1.1-2009. This Standard was revised based on YY/T 0654-2008. Compared with YY/T 0654-2008, in addition to the editorial modifications, the main technical changes are as follows. - modified the applicable scope as automatic biochemical analyzer for quantitative analysis of various samples by ultraviolet-visible spectrophotometry; - in normative references, modified GB/T 14710 “The environmental requirements and test methods for medical electrical equipment” to GB/T 14710 “Medical electrical environment requirements and test methods.’ - deleted GB/T 2829 “Sampling procedures and tables for periodic inspection by attributes (Apply to inspection of process stability)” from normative references; - deleted YY 0466 “Medical devices - Symbols to be used with medical device labels labelling and information to be supplied” (ISO 15233.2000, IDT) from normative references; - modified the sample carryover rate as not exceeding 0.1% and modified its test methods (see 5.8, 6.7); - adjusted the UREA (urea) test range in the intra-assay precision of clinical projects to 7.0 mmol/L ~ 11.0 mmol/L (see 5.10); - added the requirements and test methods for applicable provisions of GB 4793.9 and YY 0648 in safety requirements (see 5.13, 6.12); - added GB/T 18268.1, GB/T 18268.26 electromagnetic compatibility requirements and test methods (see 5.14, 6.13); - in the absorbance stability test method, the potassium dichromate solution used at a wavelength of 340 nm was changed to a yellow orange G solution (see 6.4); - in the absorbance repeatability test method, the potassium dichromate solution used at a wavelength of 340 nm was changed to a yellow orange G solution (see 6.5); - the accuracy and repeatability of the test method is modified by the manufacturer to choose one of two methods (see 6.8); Automatic chemistry analyzer 1 Scope This Standard specifies the terms and definitions, classification, requirements, test methods, marks and instructions for use, packaging, transport and storage of automatic chemistry analyzer (hereinafter referred to as the analyzer). This Standard is applicable to automatic biochemical analyzer for quantitative analysis of various samples by UV-visible spectrophotometry. 2 Normative references The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. GB/T 191, Packaging and storage marks GB 4793.1, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1. General requirements GB 4793.9, Safety requirements for electrical equipment for measurement, control and laboratory use - Part 9. Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes GB/T 14710, Environmental requirement and test methods for medical electrical equipment GB/T 18268.1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1. General requirements GB/T 18268.26, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 26. Particular requirements - In vitro diagnostic (IVD) medical equipment GB/T 29791.3, In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 3. In vitro diagnostic instruments for professional use YY 0648, Safety requirements for electrical equipment for measurement control and laboratory use - Part 2-101. Particular requirements for in vitro Recycling or single use. 5 Requirements 5.1 Normal working environment 5.1.1 Power supply voltage. 220V ± 22V, 50Hz ± 1Hz. 5.1.2 Ambient temperature. 15°C~30°C. 5.1.3 Relative humidity. 40%~85%. 5.1.4 Atmospheric pressure. 86.0kPa~106.0kPa. NOTE. If the conditions in 5.1.2~5.1.4 are inconsistent with the conditions stated in the manufacturing mark, the conditions specified in the product shall prevail. 5.2 Stray light Absorbance is not less than 2.3. 5.3 Absorbance linear range The maximum absorbance of the relative bias within ±5% shall be no less than 2.0. 5.4 Absorbance accuracy Comply with Table 1. Table 1 -- Absorbance accuracy requirements Absorbance value Allowable error 0.5 ±0.025 1.0 ±0.07 5.5 Absorbance stability The change in absorbance shall not be greater than 0.01. 5.6 Absorbance repeatability Expressed by coefficient of variation, not exceeding 1.5%. 5.7 Temperature accuracy and volatility The temperature value is within ±0.3°C of the set value, and the fluctuation is not more than ±0.2°C. 5.8 Sample carry-over rate ci - relative concentration; i - concentration serial number, range is 5~11. Where, Ai - the average value of the absorbance actually measured at a certain concentration point; ci - relative concentration; n - selected number of concentration; i - concentration serial number, range is 1~4. The absorbance range of relative bias less than ±5% is the linear range of absorbance and shall meet the requirements of 5.3. 6.3 Absorbance accuracy Use deionized water as a reference. Determine the absorbance of the potassium dichromate standard solution of which the absorbance is about 0.5 (with deionized water as blank, allowable deviation is ±5%) and 1.0 (with deionized water as blank, allowable deviation of ±5%) at 340 nm on the analyzer. Repeat the measurement 3 times and calculate the difference between the arithmetic mean of the 3 measurements and the standard value, which shall meet the requirements of Table 1. 6.4 Absorbance stability Perform the absorbance stability measurement at any wavelength in the 340 nm and 600 nm to 700 nm wavelength range of the analyzer. The 340 nm measurement solution is Orange G standard solution with an absorbance of 0.5 (with deionized water as a blank and an allowable deviation of ±5%). The measurement solution at any wavelength in the wavelength range of 600 nm to 700 nm has an absorbance of 0.5 (in terms of absorbance of 0.5). Deionized water is blank, allowing a deviation of ±5%) of copper sulfate standard solution. According to the following setting conditions a) and b), the absorbance of the above solution shall be measured on the analyzer, and the difference between Vr - reagent addition volume. NOTE 1. Surfactants can be added to deionized water (such as TritonX-100). NOTE 2. Allow adoption of 600nm~700nm as a sub-wavelength. 6.8 Sample loading accuracy and repeatability It can be divided into two types. colorimetric and weighing methods, manufacturers can choose one of two methods. 6.8.1 Weighing method The weighing method is determined as follows. a) Start the test after balancing the analyzer, degassed deionized water, etc. in a constant temperature, constant humidity laboratory for several hours. Prepare a suitable container (to prevent moisture evaporation in the container) and zero on an electronic balance with a graduation value of 0.01 mg; b) Place the container in the appropriate position, control the reagent needle or sample needle to add the specified amount of degassed deionized water to the container, and weigh the mass on the electronic balance; c) Repeat weighing 20 times for each specified addition amount. The actual amount added each time is equal to the mass of degassed deionized water divided by the density of pure water at that temperature. The density of pure water at different temperatures is shown in Annex B. Calculate the coefficient of variation according to formula (4), calculate the loading error according to formula (6), and the result shall meet the requirements of 5.9. Where, Bi - loading deviation; Xi - average amount of actual addition; T - prescribed amount. 6.8.2 Colorimetric method Colorimetric method is determined by the following steps. a) Preparation of Orange G serum (pigment solution). weigh 0.35 g of Orange G powder with an electronic balance with a graduation value of measured on a spectrophotometer (478nm ± 1nm). Calculate the actual sample addition amount according to formula (9). e) Calculate the coefficient of variation of the sample and the accuracy of the sample according to formula (4) and formula (6). The result shall meet the requirements of 5.9. 6.9 Intra-assay precision of clinical projects Repeat the test using the normal value control serum or fresh patient serum for the items and concentration ranges specified in 5.10 with the reagents, calibrators, and appropriate assay procedures specified by the manufacturer. Repeat 20 times for each item. Calculate the coefficient of variation according to formula (4), which shall meet the requirements of 5.10. 6.10 Appearance Perform visual inspection, which shall comply with the provisions of 5.11. 6.11 Environmental test Test in accordance with the method specified in GB/T 14710, the results shall comply with the provisions of 5.12. 6.12 Safety requirements Test according to the methods specified in GB 4793.1, GB 4793.9, YY 0648, the results shall meet the requirements of 5.13. 6.13 Electromagnetic compatibility Test according to the method specified in GB/T 18268.1 and GB/T 18268.26, the results shall comply with the provisions of 5.14. 7 Marks and instructions for use Meet the requirements of GB/T 29791.3. 8 Packaging, transport and storage 8.1 Packaging ......

BASIC DATA
Standard ID YY/T 0654-2017 (YY/T0654-2017)
Description (Translated English) Automatic chemistry analyzer
Sector / Industry Medical Device & Pharmaceutical Industry Standard (Recommended)
Classification of Chinese Standard C44
Classification of International Standard 11.100
Word Count Estimation 12,115
Date of Issue 2017-03-28
Date of Implementation 2018-04-01
Older Standard (superseded by this standard) YY/T 0654-2008
Drafting Organization Beijing Medical Device Inspection Institute, Hitachi High-Tech (Shanghai) International Trade Co., Ltd. Beijing Branch, Shanghai Branch, the experimental system Co., Ltd., Song Song Technology Co., Ltd. Beijing, Roche Diagnostics (Shanghai) Co., Ltd., Beckman Coulter Commerce (China) Co., Ltd.
Administrative Organization National Medical Clinical Laboratory and in vitro diagnostic system standardization Technical Committee (SAC/TC 136)
Regulation (derived from) China Food & Drug Administration Announcement 2017 No. 38
Proposing organization China Food and Drug Administration
Issuing agency(ies) State Food and Drug Administration

BASIC DATA
Standard ID YY/T 0654-2008 (YY/T0654-2008)
Description (Translated English) Automatic Chemistry Analyzer
Sector / Industry Medical Device & Pharmaceutical Industry Standard (Recommended)
Classification of Chinese Standard C44
Classification of International Standard 11.100
Word Count Estimation 13,129
Date of Issue 2008-04-25
Date of Implementation 2009-06-01
Quoted Standard GB/T 191; GB 4793.1; GB/T 2829-2002; GB/T 14710; YY 0466-2003
Drafting Organization Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Administrative Organization National Standardization Technical Committee of Medical clinical testing laboratories and in vitro diagnostic systems
Regulation (derived from) SFDA [2008] No. 192
Proposing organization National Food Quality and Safety Supervision and Inspection Center
Issuing agency(ies) State Food and Drug Administration
Summary This standard specifies the automatic biochemical analyzer (hereinafter referred to as the analyzer) the terms and definitions, classification, requirements, test methods, inspection rules, signs and instructions, packaging, transportation and storage. This standard applies to the use of liquid reagents for quantitative analysis of various samples analyzer.

BASIC DATA
Standard ID ()
Description (Translated English) (Technical specifications for ceramic capacitive sensor type partial discharge monitoring devices of 35kV and below)
Sector / Industry Chinese Industry Standard
Date of Issue 2023-10-11
Date of Implementation 2024-04-11
Issuing agency(ies) National Energy Board