HOME   Cart(0)   Quotation   About-Us Tax PDFs Standard-List Powered by Google www.ChineseStandard.net Database: 189760 (2 Nov 2024)

JJF 1059.1-2012 Related PDF English (JJF1059-1999)

JJF 1059.1-2012 (JJF1059.1-2012) & related versions
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)See DetailStatusSimilar PDF
JJF 1059.1-2012English760 Add to Cart 0-9 seconds. Auto delivery. [Including 2013XG] Evaluation and Expression of Uncertainty in Measurement JJF 1059.1-2012 Valid JJF 1059.1-2012
JJF 1059-1999EnglishRFQ ASK 9 days Evaluation and Expression of Uncertainty in Measurement JJF 1059-1999 Obsolete JJF 1059-1999
Buy with any currencies (Euro, JPY, KRW...): JJF 1059.1-2012    Preview this PDF: JJF 1059.1-2012



JJF 1059.1-2012 JJF METEOROLOGICAL INDUSTRY STANDARD OF THE PEOPLE’S REPUBLIC OF CHINA Evaluation and expression of uncertainty in measurement [Including Amendment 2013XG1] ISSUED ON: DECEMBER 03, 2012 IMPLEMENTED ON: JUNE 03, 2013 Issued by: General Administration of Quality Supervision, Inspection and Quarantine Table of Contents Introduction ... 7  1 Scope ... 10  2 Normative references ... 11  3 Terms and definitions ... 12  3.1 Measured [JJF 1001,4.7]... 12  3.2 Measurement results, result of measurement [JJF 1001,5.1] ... 13  3.3 Measured quantity value [JJF 1001,5.2] ... 13  3.4 Measurement precision [JJF 1001, 5.10] ... 14  3.5 Measurement repeatability [JJF 1001, 5.13] ... 14  3.6 Measurement repeatability condition of measurement [JJF 1001, 5.14] ... 14  3.7 Measurement reproducibility [JJF 1001, 5.16] ... 15  3.8 Measurement reproducibility condition of measurement [JJF 1001, 5.15] ... 15  3.9 Intermediate precision condition of measurement [JJF 1001, 5.11] ... 15  3.10 Experimental standard deviation [JJF 1001, 5.17] ... 16  3.11 Measurement error, error of measurement [JJF 1001, 5.3] ... 17  3.12 Measurement uncertainty, uncertainty of measurement [JJF 1001, 5.18] ... 17  3.13 Standard uncertainty [JJF 1001, 5.19] ... 18  3.14 Type A evaluation of measurement uncertainty [JJF 1001, 5.20] ... 18  3.15 Type B evaluation of measurement uncertainty [JJF 1001, 5.21] ... 18  3.16 Combined standard uncertainty [JJF 1001, 5.22] ... 19  3.17 Relative standard uncertainty [JJF 1001, 5.23] ... 19  3.18 Expanded Uncertainty [JJF 1001, 5.27] ... 19  3.19 Coverage interval [JJF 1001, 5.28] ... 20  3.20 Coverage probability [JJF 1001, 5.29]... 20  3.21 Coverage factor [JJF 1001, 5.30] ... 20  3.22 Measurement model, model of measurement [JJF 1001, 5.31] ... 21  3.23 Measurement function [JJF 1001, 5.32] ... 21  3.24 Input quantity in in a measurement model [JJF 1001, 5.33] ... 21  3.25 Output quantity in the measurement model [JJF 1001, 5.34] ... 22  3.26 Definitional uncertainty [JJF 1001, 5.24] ... 22  3.27 Instrumental measurement uncertainty [JJF 1001, 7.24] ... 22  3.28 Null measurement uncertainty [JJF 1001, 7.25] ... 23  3.29 Uncertainty budget [JJF 1001, 5.25] ... 23  3.30 Target uncertainty [JJF 1001, 5.26] ... 23  3.31 Degrees of freedom... 23  3.32 Covariance ... 24  3.33 Correlation coefficient ... 25  4 Evaluation method of measurement uncertainty ... 25  4.1 Analysis of sources of measurement uncertainty ... 26  4.2 Establishment of measurement model ... 27  4.3 Evaluation of standard uncertainty ... 30  4.4 Calculation of combined standard uncertainty ... 42  4.5 Determination of extended uncertainty ... 50  5 Report and expression of measurement uncertainty ... 51  5.1 Report of measurement uncertainty ... 51  5.2 Expression of measurement uncertainty ... 53  5.3 Other requirements when reporting uncertainty ... 55  6 Application of measurement uncertainty ... 56  6.1 Requirements for reporting measurement uncertainty in calibration certificate ... 56  6.2 Laboratory calibration and expression of measurement capability ... 57  6.3 Application in other situations ... 58  Appendix A Examples of evaluation methods of measurement uncertainty (reference) ... 59  Appendix B Table of tp(ν) values (t values) for t distributions with different probabilities p and degrees of freedom ν (supplementary) ... 92  Appendix C Summary of symbols related to quantities (supplementary) ... 94  Appendix D English-Chinese contrast of terms (reference) ... 97  Amendment No.1 to JJF 1059.1-2012 "Evaluation and expression of uncertainty in measurement" ... 99  Evaluation and expression of uncertainty in measurement 1 Scope a) The general method for evaluating and expressing measurement uncertainty specified in this specification is applicable to measurement fields of various accuracy levels, such as: 1) The establishment of national measurement standards and measurement standards at all levels and the comparison of values; 2) The setting value of standard substance and the release of standard reference data; 3) Preparation of technical documents such as measurement methods, verification procedures, verification system tables, calibration specifications, etc.; 4) Expression of measurement results and measurement capabilities in measurement qualification recognition, measurement confirmation, quality certification, laboratory accreditation; 5) Calibration, verification and other measurement services of measuring instruments; 6) Measurement in the fields of scientific research, engineering, trade settlement, medical and health care, safety protection, environmental monitoring, resource protection. b) This specification mainly concerns the measurement uncertainty of the measured estimated value that is clearly defined and can be characterized by a unique value. As for the measured quantity value that appears as a distribution of a series of values or depends on one or more parameters (for example, with time as the parameter variable), the description of the measured quantity value shall be a set of values, the distribution and relationship shall be given. c) This specification is also applicable to the evaluation and expression of uncertainties in the design and theoretical analysis of experiments, measuring methods, measuring devices, complex components and systems. 3 Terms and definitions The metrology terminology in this specification adopts JJF 1001-2011, which is based on the revision of international standard ISO/IEC GUIDE 99:2007 (the third edition of VIM). The probability and statistical terms used in this specification basically adopt the terms and definitions of the international standard ISO 3534-1:2006. 3.1 Measured [JJF 1001,4.7] The amount to be measured. Note: 1 The description of the measured requires an understanding of the type of quantity and the description of the phenomena containing the quantity, objects or substance status, including relevant components and chemical entities. 2 In the second edition of VIM and IEC 60050-300:2001, the measured is defined as the measured quantity. 3 The measurement includes the measurement system and the conditions under which the measurement is carried out. It may change the phenomenon, object substance in the study, so that the measured quantity may be different from the defined measured. In this case, it needs necessary corrections to be made. Example: 1 When measuring with a voltmeter with insufficient internal resistance, the potential difference between the two ends of the battery will decrease; the open circuit potential difference can be calculated based on the internal resistance of the battery and the voltmeter. 2 The length of the steel bar when it is in equilibrium with the ambient temperature of 23 °C is different from the length when the specified temperature to be measured is 20 °C. In this case, it must be corrected. 3 In chemistry, the name of “analyte” or substance or compound is sometimes called “measured”. This usage is wrong because these terms do not involve quantity. measurement is not too small compared with the measurement uncertainty, the measured quantity value is usually an average or median estimate of a set of true values. 4 In the Guide to Measurement Uncertainty (GUM), the terms used for the measured quantity values are "measurement results" and "estimated measured quantity value" or "estimated measured quantity value". 3.4 Measurement precision [JJF 1001, 5.10] Referred to as precision Under the specified conditions, the degree of agreement between the measured indication value and the measured quantity value through repeated measurement of the same or similar measured object. Note: 1 Measurement precision is usually expressed in numerical form with imprecision, such as standard deviation, variance or coefficient of variation under specified measurement conditions. 2 The prescribed conditions may be repetitive measurement conditions, intermediate precision measurement conditions or reproducible measurement conditions. 3 Measurement precision is used to define measurement repeatability, intermediate measurement precision or measurement repeatability. 4 The term "measurement precision" is sometimes used to refer to "measurement accuracy", which is wrong. 3.5 Measurement repeatability [JJF 1001, 5.13] Referred to as repeatability. Measurement precision under a set of repeatable measurement conditions. 3.6 Measurement repeatability condition of measurement [JJF 1001, 5.14] Referred to as repeatability condition. A set of measurement conditions of repeated measurement on the same or 3.11 Measurement error, error of measurement [JJF 1001, 5.3] Abbreviated as error. The measured quantity value minus the reference value. Note: 1 The concept of measurement error can be used in the following two situations: ① When it involves the existence of a single reference value, such as calibration with a measurement standard whose measurement uncertainty of the measured quantity value is negligible, or when a quantitative value is given, the measurement error is known; ② Assuming that the measured is characterized by a unique set of true values or a set of true values with negligible range, the measurement error is unknown. 2 Measurement errors shall not be confused with errors or faults. 3.12 Measurement uncertainty, uncertainty of measurement [JJF 1001, 5.18] Abbreviated as uncertainty According to the information used, characterize the non-negative parameters that give the measured quantity value dispersion. Note: 1 Measurement uncertainty includes components caused by the influence of the system, such as the component related to the correction value and the assigned value of the measurement standard as well as the definitional uncertainty. Sometimes the estimated system impact is not corrected, but treated as an uncertainty component. 2 This parameter can be such as the standard deviation (or a specific multiple of it) called the standard measurement uncertainty, or it can specify the half width of the interval containing the probability. 3 The measurement uncertainty generally consists of several components. Some of these components can be evaluated according to the statistical different from the type A evaluation of measurement uncertainty. Example: The evaluation is based on the following information: - The value issued by the authority; - The value of certified reference materials; - Calibration certificate; - Instrument drift; - Accuracy level of certified measuring instruments; - Inferred limit values based on personnel experience. 3.16 Combined standard uncertainty [JJF 1001, 5.22] Full name is combined standard measurement uncertainty. The standard measurement uncertainty of the output quantity obtained from the standard measurement uncertainty of each input quantity in a measurement model. Note: In the case where the input in the measurement model is relevant, the covariance must be considered when calculating the combined standard uncertainty. 3.17 Relative standard uncertainty [JJF 1001, 5.23] Full name is relative standard measurement uncertainty The absolute value that is obtained by dividing the standard uncertainty by the measured quantity value. 3.18 Expanded Uncertainty [JJF 1001, 5.27] Full name is expanded measurement uncertainty. The product of the combined standard uncertainty and a digital factor greater than 1. Note: 1 This factor depends on the type of probability distribution of the output in 3.22 Measurement model, model of measurement [JJF 1001, 5.31] Referred to as model. The mathematical relationship between all known quantities involved in the measurement. Note: 1 The general form of the measurement model is the formula: h(Y, X1,..., XN) = 0, where the output Y in the measurement model is measured, whose value is derived from the relevant information of the input quantity X1, ..., XN in the measurement model. 2 In more complex situations with two or more outputs, the measurement model contains more than one formula. 3.23 Measurement function [JJF 1001, 5.32] In the measurement model, when the value calculated from the known quantity of the input quantity is the measured quantity value of the output quantity, the functional relationship between the input quantity and the output quantity. Note: 1 If the measurement model h (Y, X1, ..., XN) = 0 can be explicitly written as Y = f(X1, ..., XN), where Y is the output in the measurement model, then the function f is the measurement function. In more layman's terms, f is an algorithm symbol, which is used to calculate the only output quantity y = f(x1, ..., xN) corresponding to the input quantity x1, ..., xN. 2 The measurement function is also used to calculate the measurement uncertainty of the measured quantity value Y. 3.24 Input quantity in in a measurement model [JJF 1001, 5.33] Referred to as input quantity. The quantity that must be measured in order to calculate the measured quantity value being measured, or its value can be obtained in other ways. B measurement uncertainty. 3 Information about the measurement uncertainty of the instrument can be given in the instrument manual. 3.28 Null measurement uncertainty [JJF 1001, 7.25] The measurement uncertainty when the measured quantity value is zero. Note: 1 The null measurement uncertainty is related to the indication of zero or close to zero. It contains the interval of the measurement being too small to know whether it can be detected, or the indication interval of the measuring instrument is caused only by noise. 2 The concept of null measurement uncertainty also applies when measuring the difference between the sample and the blank. 3.29 Uncertainty budget [JJF 1001, 5.25] Statement of measurement uncertainty, including the components of measurement uncertainty and their calculation and synthesis. Note: The uncertainty report shall include the measurement model, the estimated value, the measurement uncertainty associated with each quantity in the measurement model, the covariance, the type of probability density distribution function used, the degrees of freedom, the evaluation type of the measurement uncertainty, the coverage factors. 3.30 Target uncertainty [JJF 1001, 5.26] Full name is target measurement uncertainty. According to the intended use of the measurement results, the measurement uncertainty which is specified as the upper limit. 3.31 Degrees of freedom In the calculation of variance, the number of sums minus the limit of sums. Note: 1 Under repeatability conditions, when n independent measurements are uncertainty of the instrument; g) The inaccuracy of the standard value provided by the measurement standard or reference material; h) Inaccurate values of quoted constants or other parameters; i) Approximations and assumptions in measurement methods and measurement procedures; j) Under the same conditions, the changes of measured repeated observations. The source of measurement uncertainty must be specifically analyzed according to the actual measurement situation. In the analysis, in addition to the definitional uncertainty, it can be comprehensively considered from the aspects of measuring instruments, measuring environment, measuring personnel, measuring methods. In particular, pay attention to the sources of uncertainty that have a greater impact on the measurement results; try to avoid missing or repetition. 4.1.3 The correction is only to compensate the system error; the correction value is uncertain. When evaluating the measurement uncertainty of the corrected measured estimated value, it shall consider the uncertainty introduced by the correction. Only when the uncertainty of the correction value is small and the contribution to the combined standard uncertainty is negligible, it may not be considered. 4.1.4 Errors or sudden factors in the measurement are not a source of measurement uncertainty. In the evaluation of measurement uncertainty, the outliers in the measured quantity values shall be eliminated. The elimination of outliers shall be carried out after proper inspection of the data. Note: The judgment and processing method of outliers can be found in GB/T 4883-2008 "Statistical interpretation of data - Detection and treatment of outliers in the normal sample". 4.2 Establishment of measurement model 4.2.1 In the measurement, when the measured (i.e. output quantity) Y is determined by N other quantities X1, X2, ..., XN (i.e. input quantity) through the function f, the formula (1) is called the measurement model: Where the capital letters represent the symbol of quantity, f is the measurement is helpful to establish the measurement model when using the verification standard and control chart method to indicate that the measurement process is always in the state of statistical control. 4.2.5 If the data indicates that the measurement function fails to model the measurement process to the accuracy required for measurement, additional input must be added to the measurement model to reflect the lack of knowledge of the amount of influence. 4.2.6 The input quantity in the measurement model can be: a) The quantity directly measured by the current. These quantities and their uncertainties can be obtained by single observation, repeated observation or empirical estimation; it can include correction values for readings of measuring instruments and correction values for influence quantities such as ambient temperature, atmospheric pressure, humidity, etc. b) Quantity introduced by external sources. Such as the quantity of calibrated measurement standards or certified reference materials, as well as the reference data found in the manual. 4.2.7 When analyzing measurement uncertainty, the uncertainty of each input quantity in the measurement model is the source of the uncertainty of the output quantity. 4.2.8 This specification is mainly applicable when the measurement model is a linear function. If it is a non-linear function, Taylor series expansion shall be adopted and its high-order terms shall be ignored; the measured quantity value shall be approximated as a linear function of the input value, so as to evaluate the measurement uncertainty. If the measurement function is obviously non- linear, the evaluation of the combined standard uncertainty must include the main higher-order terms in the Taylor series expansion. 4.2.9 When the best estimated value y of the measured Y is obtained through the estimated value x1, x2, ..., xN of the input quantities X1, X2, ..., XN, there are two calculation methods: formula (5) and formula (6): a) Calculation method I: Where, y is the arithmetic average of the measured quantity values yk obtained by n independent measurements of Y; the uncertainty of each measured quantity value yk is the same; each yk is obtained by a complete set of measured quantity value based on the N input quantities Xi as obtained at the same time. This method can increase the degree of freedom of the standard uncertainty of the evaluation, thus improve the credibility. 4.3.2.6 Pre-evaluation of repeatability In the routine verification, calibration or detection work of the same type of the test piece, if the measurement system is stable and there is no obvious change in the measurement repeatability, the measurement system can be used, using the same procedure, operator, at the same location with the test piece under measurement, to make n pre-measurements on the typical measured quantity value of the typical test piece (generally n is not less than 10), to calculate the experimental standard deviation s(xk) of the single measured quantity value from the Bessel formula, that is, the measurement repeatability. When actually measuring a certain test piece, it can only measure n’ times (1 ≤ n’ < n), take the arithmetic average value of n’ independent measurements as the estimated value of the measured, then the type A standard uncertainty of this measured estimate value due to repeatability is calculated according to formula (20): The degree of freedom of standard uncertainty as evaluated by this method is still ν = n - 1. It shall be noted that when the repeatability of the measurement is suspected to be changed, the experimental standard deviation s(xk) shall be re-measured and calculated in time. 4.3.2.7 When the estimated value xi of the input quantity Xi is obtained from the curve fitted by the experimental data by the method of least squares, any point on the curve and the standard uncertainty characterizing the curve fitting parameters can be evaluated by the relevant statistical procedures. If the measured estimated value xi presents a time-dependent random change in multiple observations, a special statistical analysis method shall be used. For example, in frequency measurement, the Allen standard deviation (Allen variance) is used. 4.3.2.8 Type A evaluation methods are usually more objective than those obtained by other evaluation methods, meanwhile are statistically rigorous, but require a sufficient number of repetitions. In addition, the measured quantity values as obtained by repeated measurements in this measurement procedure shall be independent of each other. 4.3.2.9 The source of random effects shall be considered as much as possible in the type A evaluation, so that it can be reflected in the measured quantity value. Note: For example: u(xi) and u(xj) - The standard uncertainty of xi and xj. 4.4.4.3 Use appropriate methods to remove correlations a) Introduce the correlation-induced quantity into the measurement model as an independent additional input quantity For example, if the measurement model of the measured estimated value is y = f(xi, xj), when determining the measured Y, use a thermometer to determine the temperature correction value xi of the estimated value of the input Xi, meanwhile use the same thermometer to determine the temperature correction value xj of the estimated value of the other input quantity Xj, the two temperature correction values xi and xj are obviously correlated. xi = F(T), xj = G(T), that is, both xi and xj are related to the temperature; because the same thermometer is used for measurement, if the thermometer shows a large value, the correction values of both are affected at the same time, so the two input quantities xi and xj in y = f[xi(T), xj(T)] are correlated. However, as long as the temperature T is used as an independent additional input in the measurement model, that is, y = f(xi, xj, T), wherein xi, xj are the estimated values of the input quantities Xi, Xj, the input quantity T has the standard uncertainty which is not related to the above two quantities, so there is no need to introduce the covariance or correlation coefficient of xi and xj when calculating the combined standard uncertainty. b) Take effective measures to transform the input quantity For example, the uncertainty component of the calibration value in the calibration of the gauge block includes two input quantities: the temperature θs of the standard gauge block and the temperature θ of the calibrated gauge block, that is, L = f(θs, θ, ...). Since the two gauge blocks are on the same measuring device in the laboratory, the temperatures θs and θ are related. But as long as θ is transformed into θ = θs + δθ, the temperature difference δθ of the calibrated block and the standard block and the temperature θs of the standard block are used as two input quantities, at this time the two input quantities are not related, that is, the θs in L = f(θs, δθ, ...) is not related to δθ. 4.4.5 Effective degrees of freedom of combined standard uncertainty 4.4.5.1 The degree of freedom of the combined standard uncertainty uc(y) is called the effective degree of freedom, which is represented by the symbol νeff. It indicates the reliability of the evaluated uc(y). The greater the νeff, the more reliable the evaluated uc(y). 4.4.5.2 Effective degrees of freedom νeff need to be calculated in the following shall be as detailed as possible, so that users can correctly use the measurement results. Only for certain uses, if the measurement uncertainty is considered to be negligible, the measurement result can be expressed as a single measured quantity value, without reporting its measurement uncertainty. 5.1.2 Generally, when reporting the following measurement results, the combined standard uncertainty uc(y) is used; if necessary, its effective degree of freedom νeff is given: a) Basic metrology research; b) Measurement of basic physical constants; c) International comparison reproducing the units of international units (according to relevant international provisions, it may also use the extended uncertainty of k = 2). 5.1.3 In addition to the above provisions or the parties concerned agreed to the use of combined standard uncertainty, usually when reporting measurement results, they are expressed as extended uncertainty. When it comes to the measurement of industry, commerce, health and safety, if there are no special requirements, the extended uncertainty U will be reported, which is generally taken as k = 2. 5.1.4 The measurement uncertainty report generally includes the following: a) The measurement model being measured; b) Sources of uncertainty; c) The value of the standard uncertainty u(xi) of the input quantity as well as its evaluation method and evaluation process; d) Sensitivity coefficient ; e) Uncertainty component of output quantity ui(y) = ci u(xi), if necessary, give the degrees of freedom of each component νi; f) Give the covariance or correlation coefficient for all relevant input quantities; g) The combined standard uncertainty uc and its calculation process, if necessary, give the effective degree of freedom νeff; h) Extended uncertainty U or Up and its determination method; i) Report the measurement results, including the estimated value being U or Up, including the unit of measurement; c) If necessary, it may also give the relative expanded uncertainty Urel; d) It shall give the value of k for U; give p and νeff for Up. 5.2.2.1 U = kuc(y) report can be in one of the following four forms. For example, the mass of the standard weight is ms, the estimated value being measured is 100.02147 g, uc(y) = 0.35 mg, the coverage factor k = 2, U = 2 × 0.35 mg = 0.70 mg, then it is reported as follows: a) ms = 100.02147 g, U = 0.70 mg; k = 2. b) ms = (100.02147 ± 0.00070) g; k = 2. c) ms = 100.02147 (70) g; for the U value of k = 2 in parentheses, the last digit is aligned with the last digit in the previous ...... ......

BASIC DATA
Standard ID JJF 1059.1-2012 (JJF1059.1-2012)
Description (Translated English) Evaluation and Expression of Uncertainty in Measurement
Sector / Industry Metrology & Measurement Industry Standard
Classification of Chinese Standard A50
Classification of International Standard 17.020
Word Count Estimation 61,697
Older Standard (superseded by this standard) JJF 1059-1999
Quoted Standard JJF 1001-2011; GB/T 70-2008; GB 3101-1993; ISO/IEC GUIDE 98-3-2008; ISO 3534-1-2006; GB/T 4883-2008
Drafting Organization Jiangsu Province Institute of Metrology
Administrative Organization National Technical Committee of Legal Metrology management
Regulation (derived from) New version of " Evaluation and Expression of Uncertainty in Measurement Technical specifications" released; AQSIQ Announcement No. 199 of 2012
Issuing agency(ies) State Administration of Quality Supervision, Inspection and Quarantine
Summary This standard applies to the level of accuracy of measurement of various fields. Required by this specification Evaluation and Expression of Uncertainty in Measurement general method, for example: a) Basis of measurement and at all levels of national meas

BASIC DATA
Standard ID JJF 1059-1999 (JJF1059-1999)
Description (Translated English) Evaluation and Expression of Uncertainty in Measurement
Sector / Industry Metrology & Measurement Industry Standard
Classification of Chinese Standard A50
Word Count Estimation 40,482
Date of Issue 1999/6/21
Date of Implementation 1999/5/1
Older Standard (superseded by this standard) JJF 1027-1991
Regulation (derived from) New version of "Evaluation and Expression of Uncertainty in Measurement technical specifications" release; National AQSIQ Notice 2012 No. 199