HOME   Cart(0)   Quotation   About-Us Tax PDFs Standard-List Powered by Google www.ChineseStandard.net Database: 189760 (5 Oct 2024)

GB/T 6682-2008 related PDF English

GB/T 6682-2008 (GB/T6682-2008, GBT 6682-2008, GBT6682-2008) & related versions
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)See DetailStatusSimilar PDF
GB/T 6682-2008English70 Add to Cart 0-9 seconds. Auto delivery. Water for analytical laboratory use -- Specification and test methods GB/T 6682-2008 Valid GBT 6682-2008
GB/T 6682-1992English70 Add to Cart 0-9 seconds. Auto delivery. Water for analytical laboratory use--Specification and test methods GB/T 6682-1992 Obsolete GBT 6682-1992
Buy with any currencies (Euro, JPY, KRW...): GB/T 6682-2008    Preview this PDF: GB/T 6682-2008



GB/T 6682-2008: PDF in English (GBT 6682-2008)
GB/T 6682-2008 GB ICS 71.040.30 G 60 NATIONAL STANDARD OF THE PEOPLE'S REPUBLIC OF CHINA Replacing GB/T 6682-1992 Water for Analytical Laboratory Use - Specification and Test Methods ISSUED ON. MAY 15, 2008 IMPLEMENTED ON. NOVEMBER 1, 2008 Issued by. General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China; Standardization Administration of the People's Republic of China. Table of Contents Foreword ... 3  1 Scope ... 4  2 Normative references ... 4  3 Appearance ... 4  4 Level ... 5  5 Specification ... 5  6 Sampling and Storage ... 6  7 Test Methods ... 7  8 Test Report ... 11  Annex A ... 12  Annex B... 13  Foreword This Standard is modified in relation to ISO 3696.1987 "Water for Analytical Laboratory Use - Specification and Test Methods" (English edition). This Standard was modified during the adoption of ISO 3696. 1987, in consideration of the national conditions of China. Those technical differences have already been incorporated into the text and marked with perpendicular single line on the margin of relevant clauses. Comparison Between chapter and article numbers of this Standard and ISO 3696.1987 is listed in Annex A. Technical differences and reasons between this Standard and ISO 3696.1987 is listed in Annex B for reference. This Standard replaces "Water for Analytical Laboratory Use - Specification and Test Methods" GB/T 6682 - 1992. Compared with GB/T 6682 - 1992, this Standard has the following main changes. - ADD the test report (Chapter 8 of this edition). Annex C of this Standard is normative; Annex A and Annex B are informative. This Standard is proposed by China Petroleum and Chemical Industry Association. This Standard is under the jurisdiction of Chemical Reagent Subcommittee of National Technical Committee on Chemical of Standardization Administration of China (SAC/TC 63/SC 3). Drafting organization of this Standard. Sinopharm Chemical Reagent Co., Ltd. Main drafters of this Standard. Chen Haoyun, Chen Hong. This Standard was issued in 1686 for the first time and revised in 1992 for the first time. Water for Analytical Laboratory Use - Specification and Test Methods 1 Scope This Standard specifies the level, specification, sampling and storage, test methods and test report of water for analytical laboratory use. This Standard is applicable to the water for chemical analysis and inorganic trace analysis tests. Water of different levels may be selected according to the requirements of actual work. 2 Normative references The provisions in following documents become the provisions of this Standard through reference in this Standard. For dated references, the subsequent amendments (excluding corrigendum) or revisions do not apply to this Standard, however, parties who reach an agreement based on this Standard are encouraged to study if the latest versions of these documents are applicable. For undated references, the latest edition of the referenced document applies. GB/T 601 Chemical reagent - Preparations of standard volumetric solutions GB/T 602 Chemical reagent - Preparations of standard solutions for impurity (GB/T 602-2002, ISO 6353-1.1982, NEQ) GB/T 603 Chemical reagent - Preparations of reagent solutions for use in test methods (GB/T 603-2002, ISO 6353-1.1982, NEQ) GB/T 9721 Chemical reagent - General rules for the molecular absorption spectrophotometry (ultraviolet and visible) GB/T 9724 Chemical reagent - General rule for the determination of pH (GB/T 9724-2007, ISO 6353-1.1982, NEQ) GB/T 9740 Chemical reagent - General method for the determination of dry residue after evaporation (GB/T 9740-2008, ISO 6353-1.1982, NEQ) 3 Appearance Water for analytical laboratory use shall be colorless, transparent liquid through visual observation. 4 Level Raw water for analytical laboratory use shall be drinking water or water in proper purity. Water for analytical laboratory use is classified into three levels. level 1 water, level 2 water and level 3 water. 4.1 Level 1 water Level 1 water is used for the analytical test with strict requirements, including the test with requirements for particles, such as water for high performance liquid chromatography analysis. Level 1 water may be prepared by level 2 water after distilled by quartz equipment or treated by ion exchange mixed bed, and then filtered through 0.2μm microporous filtering film. 4.2 Level 2 water Level 2 water is used for inorganic trace analysis test, such as water for atomic absorption spectroscopy. Level 2 water may be prepared by such methods as several distillation or ion exchange. 4.3 Level 3 water Level 3 water is used for general chemical analysis test. Level 3 water may be prepared by such methods as distillation or ion exchange. 5 Specification Specification of water for analytical laboratory use is detailed in Table 1. 6.3 Storage During the storage period of water at various levels, the main contamination sources are soluble constituent of container, carbon dioxide in the air and other impurities. Therefore, level 1 water shall not be stored, but prepared before application. Level 2 water and level 3 water may be prepared in the right amount. They shall be respectively stored in the corresponding containers washed with water of the same level in advance. Water at various levels shall be prevented from being contaminated during the transportation process. 7 Test Methods In the test methods, each test shall be carried out in the clean environment and adequate measures shall be taken to prevent the sample from being contaminated. Water sample shall be measured out to the accuracy of 0.1mL; adopted solution represented by "%" is mass fraction. Analytical reagent and water of corresponding level shall be used in the test. 7.1 pH value 100mL of water sample shall be measured out and determined according to the requirements of GB/T 9724. 7.2 Conductivity 7.2.1 Instrument 7.2.1.1 Conductivity meter for level 1 and level 2 water determination. equipped with "online" conductivity cell with 0.01cm-1~0.1cm-1 electrode constant, have temperature automatic compensation function. In the case of conductivity meter without temperature compensation function, "online" heat exchanger may be installed to control the water temperature 25°C±1°C in the determination; or water temperature may be recorded and reduced according to Annex C. 7.2.1.2 Conductivity meter for level 3 water determination. equipped with conductivity cell with 0.1cm-1~1cm-1 electrode constant, have temperature automatic compensation function. In the case of conductivity meter without temperature compensation function, thermostatic water-bath groove may be installed to control the to-be-measured water 7.4.1 Instrument conditions Quartz absorption cell. thickness of 1cm and 2cm. 7.4.2 Determination procedures Respectively POUR the water sample into 1cm and 2cm absorption cells, USE the water sample in 1cm absorption cell as reference at 254nm, DETERMINE the absorbance of water sample in 2cm absorption cell. If the sensitivity of instrument is insufficient, the thickness of measurement absorption cell may be added properly. 7.5 Evaporation residue 7.5.1 Instrument 7.5.1.1 Rotary evaporator. equipped with 500mL distillation flask. 7.5.1.2 Thermostatic water-bath. 7.5.1.3 Evaporation pan. the materials may be platinum, quartz and borosilicate glass. 7.5.1.4 Electric oven. temperature may be controlled between 105°C±2°C. 7.5.2 Determination procedures 7.5.2.1 Water sample pre-contraction MEASURE out 1000mL of level 2 water (500mL of level 3 water), ADD the water sample into the distillation flask of rotary evaporator in several times, CARRY out reduced pressure evaporation on the water bath (avoiding evaporate to dryness), SYOP heating until the water sample is evaporated to 50mL. 7.5.2.2 Determination TRANSFER the above pre-contraction water sample INTO a evaporation pan constant volume at 105°C±2°C, RINSE the distillation flask with 5mL~10mL of water sample for twice, COMBINE the washing liquid and pre-contraction water sample into a evaporation pan, DETERMINE according to the requirements of GB/T 9740. 7.6 Soluble silicon 7.6.1 Reagent preparation 7.6.1.1 Silicon dioxide standard solution (1mg/mL) It shall be prepared according to the requirements of GB/T 602. The preparation of standard colorimetric solution. TAKE 0.50 mL of silicon dioxide standard solution (0.01mg/mL), DILUTE to 20mL with water sample, treat meanwhile and likewise with the test solution of the same v... ......

BASIC DATA
Standard ID GB/T 6682-2008 (GB/T6682-2008)
Description (Translated English) Water for analytical laboratory use. Specification and test methods
Sector / Industry National Standard (Recommended)
Classification of Chinese Standard G60
Classification of International Standard 71.040.30
Word Count Estimation 11,144
Date of Issue 2008-05-15
Date of Implementation 2008-11-01
Older Standard (superseded by this standard) GB/T 6682-1992
Quoted Standard GB/T 601; GB/T 602; GB/T 603; GB/T 9721; GB/T 9740; GB/T 9724
Adopted Standard ISO 3696-1987, MOD
Drafting Organization Sinopharm Chemical Reagent Co., Ltd.
Administrative Organization National Chemical Standardization Technical Committee
Regulation (derived from) Announcement of Newly Approved National Standards No. 8, 2008 (No. 121 overall)
Proposing organization China Petroleum and Chemical Industry Association
Issuing agency(ies) Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China; Standardization Administration of China
Summary This standard specifies the laboratory analysis of water levels, specifications, sampling and storage, test methods and test reports, This standard applies to chemical analysis and inorganic trace analysis, test water, according to the actual work needs to choose different levels of water,