YY/T 1469-2016 PDF English
US$240.00 · In stock · Download in 9 secondsYY/T 1469-2016: Electrical infusion pump for ambulatory use Delivery: 9 seconds. True-PDF full-copy in English & invoice will be downloaded + auto-delivered via email. See step-by-step procedureStatus: Valid
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivery | Name of Chinese Standard | Status |
| YY/T 1469-2016 | English | 240 |
Add to Cart
|
0-9 seconds. Auto-delivery
|
Electrical infusion pump for ambulatory use
| Valid |
Excerpted PDFs (Download full copy in 9 seconds upon purchase)PDF Preview: YY/T 1469-2016
YY/T 1469-2016: Electrical infusion pump for ambulatory use---This is an excerpt. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.), auto-downloaded/delivered in 9 seconds, can be purchased online: https://www.ChineseStandard.net/PDF.aspx/YYT1469-2016
YY
PHARMACEUTICAL INDUSTRY STANDARD
ICS 11.040.20
C 31
Electrical infusion pump for ambulatory use
Issued on. JANUARY 26, 2016
Implemented on. JANUARY 01, 2017
Issued by. China Food and Drug Administration
Table of Contents
Foreword... 3
1 Scope... 4
2 Normative references... 4
3 Terms and definitions... 5
4 Technical requirements... 6
5 Test method... 11
Appendix A (Informative) Guidelines and compilation instructions... 20
Foreword
This Standard was drafted in accordance with the rules given in GB/T 1.1-2009.
Please note that some of the contents of this document may involve patents.
The issuing organization of this document is not responsible for identifying
these patents.
This Standard was proposed by China Food and Drug Administration.
This Standard shall be under the jurisdiction of Subcommittee 5 on
Electromedical Equipment of National Technical Committee 10 on Medical
Electrical Equipment of Standardization Administration of China (SAC/TC
10/SC 5).
The drafting organizations of this Standard. Jiangsu Medical Device Inspection
Institute, Shanghai Medical Device Testing Institute, Nantong Apon Medical
Appliance Co., Ltd.
The drafters of this Standard. Li Ning, Xu Jin, Wu Jiajun, Zhang Yichuan, Zhuo
Yue, Chen Tao, Miao Jia.
Electrical infusion pump for ambulatory use
1 Scope
This Standard specifies the definition, basic requirements and corresponding
test methods of electrical infusion pump for ambulatory use (hereinafter referred
to as infusion pump for ambulatory use).
This Standard applies to electrical infusion pump for ambulatory use which are
used in medical monitoring environments; the intended use of this product is for
intravenous or epidural analgesic administration.
The infusion pump for ambulatory use is mainly composed of a drive unit, a
restoration set and an infusion line; the restoration set and the infusion line are
single-use components.
This Standard does not apply to the following equipment.
-- equipment which is specially used for diagnostics or similar purposes (such
as high-pressure injectors);
-- gastrointestinal feeding pump;
-- equipment for extracorporeal blood circulation;
-- insulin pump and pumps for similar clinical applications;
-- infusion pump for ambulatory use that has special requirements for
infusion accuracy.
2 Normative references
The following documents are indispensable for the application of this document.
For dated references, only the dated version applies to this document. For
undated references, the latest edition (including all amendments) applies to this
document.
GB/T 1962.1-2015, Conical fittings with a 6% (Luer) taper for syringes,
needles and certain other medical equipment - Part 1.General requirement
GB/T 1962.2-2001, Conical fittings with a 6% (Luer) taper for syringes,
needles and certain other medical equipment - Part 2.Lock fittings
GB 9706.1-2007, Medical electrical equipment - Part 1.General
requirements for safety
GB 9706.27-2005, Medical electrical equipment - Part 2-24.Particular
requirements for the safety of infusion pumps and controllers
GB/T 14233.1-2008, Test methods for infusion, transfusion, injection
equipment for medical use - Part 1.Chemical analysis methods
GB/T 14233.2-2005, Test methods for infusion, transfusion, injection
equipment for medical use - Part 2.Biological test methods
GB/T 14710-2009, Environmental requirement and test methods for medical
electrical equipment
3 Terms and definitions
Terms and definitions determined by GB 9706.27-2005 and the following ones
are applicable to this document. For ease of use, some of the terms and
definitions in GB 9706.27-2005 are repeated below.
3.1 Infusion pump
The device that is expected to control the flow of fluid into the patient's body by
the positive pressure that is generated by the pump.
3.2 Infusion pump for ambulatory use
The device that is used to control patient infusion and can be continuously
carried by the patient.
3.3 Electrical infusion pump for ambulatory use
The infusion pump for ambulatory use that is driven by electrical energy.
3.4 Drive unit
The device that transmits the liquid to the patient through the infusion line by
the positive pressure that is generated by the motor drive.
3.5 Restoration set
The device for storing liquid in the infusion pump for ambulatory use.
3.6 Administration set
The device that transmits the fluid from the fluid supply site to the patient via
the device.
4 Technical requirements
4.1 Overall performance requirements
4.1.3 Infusion accuracy
4.1.3.1 The accuracy of the average infusion speed in continuous infusion
mode shall be better than ±10%.
4.1.4 Blocking pressure, triggering time and the produced bolus volume
4.1.4.1 The infusion blocking pressure alarm threshold value shall be specified
in the random file; the accuracy shall be better than ±20%.
4.2 Alarm information requirements
4.2.1 The infusion pump for ambulatory use shall contain at least the following
alarm information, and shall use visual alarm and auditory (or other effective
methods such as vibration) alarm.
4.2.2 The prompt sound pressure level of the infusion pump for ambulatory use
shall not be lower than 45 dB(A).
4.3 Requirements for the single-use part
4.3.1 Physical properties
4.3.1.5 Restoration set
The restoration set of the infusion pump for ambulatory use shall be designed
to be able to visually inspect the foreign matter and air bubbles in the contained
solution.
4.3.1.6 Particulate pollution
When the test is performed in accordance with Appendix A of YY 0451-2010 or
equivalent method, the pollution index shall not be greater than 90.
4.3.2 Chemical properties
4.3.2.1 Reducing substances
The volume difference of potassium permanganate solution [c(KMnO4) = 0.002
mol/L] that is consumed between the test solution and the blank solution shall
not exceed 2.0 mL.
4.3.2.4 Evaporation residue
The total amount of evaporation residue shall not exceed 2 mg.
4.3.2.5 UV absorbance
When the test is performed in the range of 220 nm ~ 360 nm, the UV
absorbance of the test solution shall not be greater than 0.3.
4.3.2.6 Residue of ethylene oxide
The residue of ethylene oxide shall be no more than 10 μg/g.
4.4 Safety requirements
4.4.1 The general safety requirements of infusion pump for ambulatory use shall
meet the requirements in GB 9706.1-2007.
4.4.2 The general requirements for electromagnetic compatibility of the infusion
pump for ambulatory use shall comply with the provisions of YY 0505-2012.
4.4.3 The alarm system of the infusion pump for ambulatory use shall comply
with the provisions in YY 0709-2009, where Table 204 in Y Y0709-2009 is
replaced by Table 1.
4.4.4 The special safety requirements of the infusion pump for ambulatory use
shall comply with the requirements in GB 9706.27-2005.
5 Test method
5.1 Test conditions
5.1.1 Temperature during the test. 5 °C ~ 40 °C.
5.1.2 Relative humidity during the test. 20% ~ 90%.
5.2 Overall performance test
5.2.3 Infusion accuracy test
5.2.3.1 Continuous infusion mode accuracy test procedures
Build the test platform as shown in Figure 1; use the grade-3 water for analysis
laboratory; install a non-infused line for testing.
5.2.3.2 PCA (bolus) infusion mode infusion volume and infusion speed
accuracy test
5.2.3.3 Chronological infusion accuracy test
Use the test device in Figure 1; use the grade-3 water for analysis laboratory;
install an unused infusion line for testing. Set the time period t1,..., ti, the
infusion speed r1, r2,..., ri; calculate the infusion accuracy error in each time
period.
5.2.4 Blocking pressure, triggering time and bolus volume test
Connect the pressure sensor to the line and build the platform according to
Figure 2.
5.4.1.8 Residual amount of liquid medicine
The residual amount of liquid medicine in the restoration set and restoration line
shall be the volume of the restoration minus the nominal volume; the result shall
meet the requirements of 4.3.1.8.
5.4.2 Chemical properties
The preparation of the test solution of 5.4.2.1 ~ 5.4.2.5 is in accordance with
the method of "Serial No. 5" in Table 1 of GB/T 14233.1-2008.
5.4.2.1 Reducing substances
Perform the test according to the provisions in 5.2.2 of GB/T 14233.1-2008,
which shall meet the requirements of 4.3.2.1.
5.4.2.6 Residue of ethylene oxide
Take samples from the restoration set; perform the test according to the
provisions in GB/T 14233.1-2008, which shall meet the requirements of 4.3.2.6.
5.4.3 Sterile and pyrogen-free
Perform the test according to the provisions in GB/T 14233.2-2005.
5.4.4 Biological evaluation
Perform the biological evaluation according to relevant parts of GB/T 16886.
5.5 Safety test
5.5.1 The general safety requirements of infusion pump for ambulatory use shall
5.6 Environmental test requirements
The environmental requirements of the infusion pump for ambulatory use shall
be tested in accordance with the climatic environment group II and mechanical
environment group II in GB/T 14710-2009 and the regulations of the
manufacturer.
...... Source: Above contents are excerpted from the full-copy PDF -- translated/reviewed by: www.ChineseStandard.net / Wayne Zheng et al.
Tips & Frequently Asked QuestionsQuestion 1: How long will the true-PDF of English version of YY/T 1469-2016 be delivered?Answer: The full copy PDF of English version of YY/T 1469-2016 can be downloaded in 9 seconds, and it will also be emailed to you in 9 seconds (double mechanisms to ensure the delivery reliably), with PDF-invoice. Question 2: Can I share the purchased PDF of YY/T 1469-2016_English with my colleagues?Answer: Yes. The purchased PDF of YY/T 1469-2016_English will be deemed to be sold to your employer/organization who actually paid for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. www.ChineseStandard.us -- YY/T 1469-2016 -- Click this link and select your country/currency to pay, the exact amount in your currency will be printed on the invoice. Full PDF will also be downloaded/emailed in 9 seconds.
How to buy and download a true PDF of English version of YY/T 1469-2016?A step-by-step guide to download PDF of YY/T 1469-2016_EnglishStep 1: Visit website https://www.ChineseStandard.net (Pay in USD), or https://www.ChineseStandard.us (Pay in any currencies such as Euro, KRW, JPY, AUD). Step 2: Search keyword "YY/T 1469-2016". Step 3: Click "Add to Cart". If multiple PDFs are required, repeat steps 2 and 3 to add up to 12 PDFs to cart. Step 4: Select payment option (Via payment agents Stripe or PayPal). Step 5: Customize Tax Invoice -- Fill up your email etc. Step 6: Click "Checkout". Step 7: Make payment by credit card, PayPal, Google Pay etc. After the payment is completed and in 9 seconds, you will receive 2 emails attached with the purchased PDFs and PDF-invoice, respectively. Step 8: Optional -- Go to download PDF. Step 9: Optional -- Click Open/Download PDF to download PDFs and invoice. See screenshots for above steps: Steps 1~3 Steps 4~6 Step 7 Step 8 Step 9
|