YY/T 1182-2020

Chinese Standard: 'YY/T 1182-2020'
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)StatusRelated Standard
YY/T 1182-2020English180 Add to Cart 0--10 minutes. Auto immediate delivery. Nucleic acids amplification test reagents(kits) Valid YY/T 1182-2020
YY/T 1182-2020Chinese17 Add to Cart <=1-day [PDF from Chinese Authority, or Standard Committee, or Publishing House]  

  In 0~10 minutes time, full copy of this English-PDF will be auto-immediately delivered to your email. See samples for translation quality.  

BASIC DATA
Standard ID YYT1182-2020 (YYT1182-2020)
Description (Translated English) Nucleic acids amplification test reagents(kits)
Sector / Industry Medical Device & Pharmaceutical Industry Standard
Classification of Chinese Standard C44
Classification of International Standard 11.100
Word Count Estimation 10,127
Date of Issue 2020-02-21
Date of Implementation 2021-01-01
Older Standard (superseded by this standard) YY/T 1182-2010
Drafting Organization Beijing Medical Device Inspection Institute, China Food and Drug Administration, Sun Yat-Sen University Daan Gene Co., Ltd., Hunan Shengxiang Biotechnology Co., Ltd., Shanghai Yuanqi Biomedical Technology Co., Ltd.
Administrative Organization National Technical Committee for Standardization of Medical Clinical Laboratory and In Vitro Diagnostic System (SAC / TC 136)
Regulation (derived from) Announcement No. 20 of 2020 by the State Drug Administration
Proposing organization State Drug Administration
Issuing agency(ies) State Drug Administration

YY/T 1182-2020
PHARMACEUTICAL INDUSTRY STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
ICS 11.100
C 44
Replacing YY/T 1182-2010
Nucleic acids amplification test reagents (kits)
核酸扩增检测用试剂(盒)
ISSUED ON: FEBRUARY 21, 2020
IMPLEMENTED ON: JANUARY 01, 2021
Issued by: National Medical Products Administration
Table of Contents
Foreword ... 3 
1 Scope ... 5 
2 Normative references ... 5 
3 Terms and definitions ... 5 
4 Classification ... 6 
5 Technical requirements ... 7 
6 Test method ... 12 
7 Labels and instructions ... 15 
8 Packaging, transportation and storage ... 15 
References ... 16 
Nucleic acids amplification test reagents (kits)
1 Scope
This Standard specifies the terms and definitions, classification technical
requirements, test methods, labels and instructions, packaging, transportation
and storage of nucleic acids amplification test reagents (kits) [hereinafter
referred to as "reagents (kits)"].
This Standard applies to the quality control of nucleic acids amplification test
reagents (kits). Nucleic acids amplification methods include polymerase chain
reaction (PCR) technology and isothermal nucleic acid amplification technology.
This Standard does not apply to the following products:
a) reagents (kits) for blood source screening;
b) reagents (kits) for gene sequencing.
This Standard is a general standard for nucleic acids amplification test reagents
(kits). For products or reagents (kits) with special standards, corresponding
product standards or technical requirements should be formulated according to
product characteristics and special standard requirements.
2 Normative references
The following documents are indispensable for the application of this document.
For dated references, only the dated version applies to this document. For
undated references, the latest edition (including all amendments) applies to this
document.
GB/T 21415-2008, In vitro diagnostic medical devices - Measurement of
quantities in biological samples - Metrological traceability of values assigned
to calibrators and control materials
GB/T 29791.2, In vitro diagnostic medical devices - Information supplied by
the manufacturer (labelling) - Part 2: In vitro diagnostic reagents for
professional use
3 Terms and definitions
The following terms and definitions are applicable to this document.
b) According to the principle of nucleic acid amplification, it can be divided
into: variable temperature amplification detection kit and isothermal
amplification detection kit;
c) Classify according to methodologies; examples of common methodologies:
PCR-fluorescent probe method, PCR-fluorescent dye method, PCR-
melting curve method, multiple fluorescence PCR method, RT-PCR
fluorescence method, nested-PCR method, gap-PCR method, loop-
mediated constant temperature amplification method.
5 Technical requirements
5.1 General requirements
5.1.1 Appearance
The reagents (kits) shall meet the appearance requirements that are specified
by the manufacturer; the appearance requirements include but are not limited
to the following: the reagents (kits) shall have complete components; the
package shall be clean, free from leakage and damage; the signs and labels
shall be clearly written.
5.1.2 Nucleic acid extraction function
The nucleic acid extraction function shall meet the following requirements:
a) For reagents (kits) that contain nucleic acid extraction components, the
manufacturer shall make appropriate requirements for nucleic acid
extraction, and verify the nucleic acid extraction function. For example:
fully consider the interference factors in the sample extraction process and
the possible impact on the subsequent sample amplification process;
b) For reagents (kits) for which samples need to be extracted, but do not
contain nucleic acid extraction components, the manufacturer shall state
or specify the extraction reagents (kits) and provide verification or
confirmation information;
c) For reagents (kits) which are directly amplified without sample extraction,
the manufacturer shall be able to provide sufficient evidence to prove the
anti-interference of the enzymes in its products.
5.1.3 Internal standard and (or) control
Manufacturers shall establish appropriate quality control procedures for the
detection results of reagents (kits); they should reasonably set internal standard
and (or) control in the reaction system according to the characteristics of their
5.2.6 Negative reference conformance rate
When it is applicable, perform determination of the negative reference in the
national reference or enterprise reference; the detection result shall be positive.
The reference settings follow the following principles:
a) Contain a certain number of samples without test substance;
b) Fully consider factors such as possible interference, crossover, non-
specific. For example: containing a certain number of other samples that
are similar to the tested species, of the same infection site, or similar
infection symptoms;
c) When it is applicable, include a certain number of genotypes that are not
within the declared detection range of the reagent (kit).
5.2.7 Detection limit
The manufacturer shall specify and provide the detection limit of the reagent
(kit); the establishing and verifying method of the detection limit of the reagent
(kit) shall be scientific and feasible. Perform determination of the detection limit
reference of the national reference or enterprise reference; the detection result
shall meet the requirements of the national reference or enterprise reference.
5.2.8 Precision
The precision within the batch shall meet that the coefficient of variation (CV, %)
of the detected concentration log value is ≤ 5%. When the sample concentration
is low, the coefficient of variation (CV, %) of the detected concentration log value
is ≤10%.
5.2.9 Stability
The following methods can be selected for verification:
a) Stability of validity period: the manufacturer shall specify the validity period
of the product. Under the storage conditions that are stipulated by the
manufacturer, take samples of the nearest validity period to test 5.1, 5.2.2
~ 5.2.8; the results shall meet the corresponding requirements;
Note 1: generally, when the validity period is 1 year, select products no
more than 1 month; when the validity period is half a year, select
products no more than half a month, and so on. However, if it exceeds
the prescribed time, the product can also be adopted when it meets the
requirements.
c) Genotyping or mutant products shall contain a certain number of
genotypes that are not within the detection range of the reagent (kit).
5.3.3 Detection limit
The manufacturer shall specify and provide the detection limit of the reagent
(kit); the establishing and verifying method of the detection limit of the reagent
(kit) shall be scientific and feasible. Perform a determination of the detection
limit reference of the national reference or enterprise reference; the detection
result shall meet the requirements of the national reference or enterprise
reference.
5.3.4 Precision
The precision within the batch shall meet the following requirements:
a) The coefficient of variation of Ct value (CV, %) ≤5%. Or
b) Test the same sample 10 times in parallel; the test results shall be
consistent.
5.3.5 Stability
The following methods can be selected to verify:
a) Stability of validity period: the manufacturer shall specify the validity period
of the product. Under the storage conditions that are stipulated by the
manufacturer, take samples of the nearest validity period to test 5.1, 5.3.1
~ 5.3.4; the results shall meet the corresponding requirements.
Note 4: generally, when the validity period is 1 year, select products no
more than 1 month; when the validity period is half a year, select products
no more than half a month, and so on. However, if it exceeds the
prescribed time, the product can also be adopted when it meets the
requirements.
b) Thermal stability test: under the thermal stability test conditions that are
specified by the manufacturer, test 5.1, 5.3.1 ~ 5.3.4; the results shall meet
the corresponding requirements.
Note 5: thermal stability cannot be directly used to derive the product
expiration date, unless a derivation formula that is based on a large
amount of stability research data is used.
Note 6: according to product characteristics, any combination of a), b)
methods can be selected; but the selected method should be able to verify
the stability of the product, so as to ensure that the product performance
within the validity period meets the standard requirements.
6.2.8 Precision
Operate according to the instructions of the reagent (kit); use at least two
samples of high and low concentration levels to repeat the test 10 times; the
concentration log value coefficient of variation (CV, %) shall meet the
requirements of 5.2.8.
6.2.9 Stability
The following methods can be selected for verification:
a) Stability of the validity period: take samples of the near validity period for
testing, which shall meet the requirements of 5.2.9a);
b) Thermal stability test: take samples within the validity period for testing
according to the method that is specified by the manufacturer, which shall
meet the requirements of 5.2.9b).
6.3 Qualitative reagents
6.3.1 Positive reference conformance rate
Operate according to the instructions of the reagent (kit); perform a detection of
the positive reference in the national reference or enterprise reference; the
d......