YY/T 0891-2013: Particular specifications for angiographic injector

---This is an excerpt. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.), auto-downloaded/delivered in 9 seconds, can be purchased online: https://www.ChineseStandard.net/PDF.aspx/YYT0891-2013
YY PHARMACEUTICAL INDUSTRY STANDARD ICS 11.040.50 C 43 Particular specifications for angiographic injector Issued on. OCTOBER 21, 2013 Implemented on. OCTOBER 1, 2014 Issued by. China Food and Drug Administration

Table of Contents

Foreword... 3 1 Scope... 4 2 Normative references... 4 3 Terms and definitions... 4 4 Composition... 6 5 Requirements... 6 6 Test methods... 8

Foreword

This Standard is drafted in accordance with the rules given in GB/T 1.1-2009. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. The issuing organization of this document shall not be held responsible for identifying any or all such patent rights. This Standard is proposed by China Food and Drug Administration. This Standard is under the jurisdiction of Subcommittee on Medical X-ray Equipment and Appliances of National Technical Committee on Medical Electrical Equipment of Standardization of Administration of China (SAC/TC 10/SC 1). Drafting organizations of this Standard. Liaoning Medical Device Test Institute, Shenzhen Seacrown Electromechanical Co., Ltd., Shenzhen Anke High-tech Co., Ltd. Drafters of this Standard. Jin Yubo, Zhong Dingsu, Yin Haibo. Particular specifications for angiographic injector

1 Scope

This Standard specifies terms and definitions, composition, requirements and test methods for angiographic injectors (hereinafter referred to as “injectors”). This Standard applies to angiography injectors.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the dated edition cited applies. For undated references, the latest edition of the referenced document (including all amendments) applies. GB 9706.1-2007 Medical electrical equipment - Part 1.General requirements for safety GB 9706.15-2008 Medical electrical equipment - Part 1.General requirements for safety - 1.Collateral standard. Safety requirements for medical electrical systems

3 Terms and definitions

For the purpose of this document, the following terms and definitions and those defined in GB/T 10149 apply. 3.1 angiographic injector The injector used to rapidly inject a sufficient amount of contrast agent to the inspection site in a certain period of time, in the diagnosis or treatment of vascular intervention angiography, in order to obtain the required medical image with the use of X-ray angiography equipment. 3.2 injector head The part equipped with injection push mechanism and supporting syringe. 3.3 flow rate The dosage of the contrast agent injected per unit of time, in milliliters per second (mL/s). 3.4 rise time The time taken from the start of the injection to when it reaches the set flow rate, in seconds. 3.5 maximum injection pressure The maximum pressure that can be reached at the time of injection of an injector.

4 Composition

The injector consists of at least a push and control system, a supporting and fixing part, accessories, etc.

5 Requirements

5.1 Working conditions 5.1.1 Environmental conditions Unless otherwise specified, the working conditions of the injector shall meet. 5.1.2 Power supply conditions The power supply conditions of the injector shall meet. 5.2 Flow rate The flow rate range of injectors for hearts and large vessels shall not be less than 1.0 mL/s to 30 mL/s. When the flow rate is less than or equal to 15 mL/s, the deviation shall not be greater than ±5 %; when the flow rate is greater than 15 mL/s, the deviation shall not be greater than ±10 %. 5.4 Fill rate The fill rate of injectors shall be specified in the product standard, the deviation shall not be greater than ±5 %. 5.5 Maximum injection pressure The maximum injection pressure of injectors for hearts and large vessels shall not be less than 6.89 MPa (1000 psi). 5.6 Pressure limit The device shall be capable of setting the pressure limit during injection. When the pressure limit is less than or equal to 4.13 MPa (600 psi), the deviation shall not exceed ±0.34 MPa (50 psi); when the pressure limit is greater than 4.13 MPa (600 psi), the deviation shall not exceed ±0.48 MPa (70 psi). 5.9 Rise time The setting range of the rise time should be specified in the product standard. 5.10 Rotation angle of injector head The angle between the upward injector head and the horizontal plane shall not be less than 60°, and the angle between the downward injector head and the horizontal plane shall not be less than 10°. 5.11 Functions The injector shall have the following functions. 5.12 Appearance It shall meet the following requirements. 5.14 Security It shall meet the requirements of GB 9706.1-2007 and GB 9706.15-2008.

6 Test methods

6.1 Working conditions 6.1.1 Environmental conditions It shall comply with the specifications of 5.1.1. 6.1.2 Power supply conditions The power supply conditions are as follows. 6.2 Flow rate The test medium is distilled or deionized water. The injection dose is set to 100 mL, and the flow rate is set to the minimum value and 50 % of the maximum value respectively. 6.3 Injection dose The test medium is distilled or deionized water. SET the injection dose to 2 mL (when the flow rate is the minimum value), 20 mL (when the flow rate is 50 % of the maximum value) and 40 mL (when the flow rate is the maximum value); USE a balance to weigh the distilled water and convert it to capacity; CALCULATE the injection dose deviation, which shall meet the requirements of 5.3. 6.4 Fill rate The test medium is distilled or deionized water. USE a timer to count the time taken from the start to the end of the filling signal; DETERMINE the capacity of the filled injection agent by weighing the weight of the container before and after filling; CALCULATE the fill rate, which shall meet the requirements of 5.4. 6.6 Pressure limit SET the pressure limit to 20 % of the maximum injection pressure (or near value) and 80 % of the maximum injection pressure (or near value) respectively; USE a tee connector to connect the syringe, the pressure gauge and the catheter; INJECT in a flow rate of 3 mL/s; READ the pressure the injector indicates when it exceeds the pressure limit and stops the injection; CALCULATE the deviation between the actual pressure and the set pressure, which shall meet the requirements of 5.5. 6.7 Injection delay SET the delay time to 3 s; USE a timer to count the time taken from receiving the signal to starting the injection; CALCULATE the injection delay deviation, which shall meet the requirements of 5.7. 6.9 Rise time Actual operation, which shall meet the requirements of 5.9. 6.10 Rotation angle of injector head Actual operation, measured with a goniometer. ......
Source: Above contents are excerpted from the full-copy PDF -- translated/reviewed by: www.ChineseStandard.net / Wayne Zheng et al.


      

Tips & Frequently Asked Questions

Question 1: How long will the true-PDF of English version of YY/T 0891-2013 be delivered?

Answer: The full copy PDF of English version of YY/T 0891-2013 can be downloaded in 9 seconds, and it will also be emailed to you in 9 seconds (double mechanisms to ensure the delivery reliably), with PDF-invoice.

Question 2: Can I share the purchased PDF of YY/T 0891-2013_English with my colleagues?

Answer: Yes. The purchased PDF of YY/T 0891-2013_English will be deemed to be sold to your employer/organization who actually paid for it, including your colleagues and your employer's intranet.

Question 3: Does the price include tax/VAT?

Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countries

Question 4: Do you accept my currency other than USD?

Answer: Yes. www.ChineseStandard.us -- YY/T 0891-2013 -- Click this link and select your country/currency to pay, the exact amount in your currency will be printed on the invoice. Full PDF will also be downloaded/emailed in 9 seconds.

How to buy and download a true PDF of English version of YY/T 0891-2013?

A step-by-step guide to download PDF of YY/T 0891-2013_EnglishStep 1: Visit website https://www.ChineseStandard.net (Pay in USD), or https://www.ChineseStandard.us (Pay in any currencies such as Euro, KRW, JPY, AUD).
Step 2: Search keyword "YY/T 0891-2013".
Step 3: Click "Add to Cart". If multiple PDFs are required, repeat steps 2 and 3 to add up to 12 PDFs to cart.
Step 4: Select payment option (Via payment agents Stripe or PayPal).
Step 5: Customize Tax Invoice -- Fill up your email etc.
Step 6: Click "Checkout".
Step 7: Make payment by credit card, PayPal, Google Pay etc. After the payment is completed and in 9 seconds, you will receive 2 emails attached with the purchased PDFs and PDF-invoice, respectively.
Step 8: Optional -- Go to download PDF.
Step 9: Optional -- Click Open/Download PDF to download PDFs and invoice.
See screenshots for above steps: Steps 1~3    Steps 4~6    Step 7    Step 8    Step 9