HOME   Cart(0)   Quotation   About-Us Policy PDFs Standard-List
www.ChineseStandard.net Database: 189759 (12 Oct 2025)

YY/T 0654-2017 PDF English

US$145.00 · In stock · Download in 9 seconds
YY/T 0654-2017: Automatic chemistry analyzer
Delivery: 9 seconds. True-PDF full-copy in English & invoice will be downloaded + auto-delivered via email. See step-by-step procedure
Status: Valid

YY/T 0654: Evolution and historical versions

Standard IDContents [version]USDSTEP2[PDF] deliveryName of Chinese StandardStatus
YY/T 0654-2017English145 Add to Cart 0-9 seconds. Auto-delivery Automatic chemistry analyzer Valid
YY/T 0654-2008English519 Add to Cart 3 days Automatic Chemistry Analyzer Obsolete

Excerpted PDFs (Download full copy in 9 seconds upon purchase)

PDF Preview: YY/T 0654-2017
      

Similar standards

YY/T 0653   YY/T 0657   YY/T 0659   YY/T 0655   

YY/T 0654-2017: Automatic chemistry analyzer

---This is an excerpt. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.), auto-downloaded/delivered in 9 seconds, can be purchased online: https://www.ChineseStandard.net/PDF.aspx/YYT0654-2017
YY PHARMACEUTICAL INDUSTRY STANDARD ICS 11.100 C 44 Replacing YY/T 0654-2008 Automatic chemistry analyzer Issued on. MARCH 28, 2017 Implemented on. APRIL 1, 2018 Issued by. China Food and Drug Administration of China

Table of Contents

Foreword... 3 1 Scope... 5 2 Normative references... 5 3 Terms and definitions... 6 4 Classification... 6 5 Requirements... 7 6 Test methods... 9 7 Marks and instructions for use... 15 8 Packaging, transport and storage... 15 Annex A (normative) Preparation method of 50g/L sodium nitrite solution... 17 Annex B (informative) Density of pure water at different temperatures at standard atmospheric pressure... 18

Foreword

This Standard was drafted in accordance with the rules given in GB/T 1.1-2009. This Standard was revised based on YY/T 0654-2008.Compared with YY/T 0654-2008, in addition to the editorial modifications, the main technical changes are as follows. - modified the applicable scope as automatic biochemical analyzer for quantitative analysis of various samples by ultraviolet-visible spectrophotometry; - in normative references, modified GB/T 14710 “The environmental requirements and test methods for medical electrical equipment” to GB/T 14710 “Medical electrical environment requirements and test methods.’ - deleted GB/T 2829 “Sampling procedures and tables for periodic inspection by attributes (Apply to inspection of process stability)” from normative references; - deleted YY 0466 “Medical devices - Symbols to be used with medical device labels labelling and information to be supplied” (ISO 15233.2000, IDT) from normative references; - modified the sample carryover rate as not exceeding 0.1% and modified its test methods (see 5.8, 6.7); - adjusted the UREA (urea) test range in the intra-assay precision of clinical projects to 7.0 mmol/L ~ 11.0 mmol/L (see 5.10); - added the requirements and test methods for applicable provisions of GB 4793.9 and YY 0648 in safety requirements (see 5.13, 6.12); - added GB/T 18268.1, GB/T 18268.26 electromagnetic compatibility requirements and test methods (see 5.14, 6.13); - in the absorbance stability test method, the potassium dichromate solution used at a wavelength of 340 nm was changed to a yellow orange G solution (see 6.4); - in the absorbance repeatability test method, the potassium dichromate solution used at a wavelength of 340 nm was changed to a yellow orange G solution (see 6.5); - the accuracy and repeatability of the test method is modified by the manufacturer to choose one of two methods (see 6.8); - the mark and instructions for use shall be changed to meet the requirements of GB/T 29791.3 (see 7); - Annex B was replaced by the 1990 International Temperature Standard Pure Water Density Table. Attention is drawn to the possibility that some of the elements of this Standard may be the subject of patent rights. The issuing authority shall not be held responsible for identifying any or all such patent rights. This Standard was proposed by China Food and Drug Administration of China. This Standard shall be under the jurisdiction of National Technical Committee on Medical Clinical Laboratory and In Vitro Diagnostic System of Standardization Administration of China (SAC/TC 136). The drafting organizations of this Standard. Beijing Medical Device Inspection Institute, Hitachi High-Tech (Shanghai) International Trade Co., Ltd. Beijing Branch, Shanghai Kehua Experimental System Co., Ltd., Beijing Songshang Technology Co., Ltd., Roche Diagnostics Products (Shanghai) Co., Ltd., Beckman Coulter Trading (China) Co., Ltd. Main drafters of this Standard. Zhao Bingfeng, Cheng Qing, Su Tao, Fu Yuguang, Tian Wei, Bi Xiao. Automatic chemistry analyzer

1 Scope

This Standard specifies the terms and definitions, classification, requirements, test methods, marks and instructions for use, packaging, transport and storage of automatic chemistry analyzer (hereinafter referred to as the analyzer). This Standard is applicable to automatic biochemical analyzer for quantitative analysis of various samples by UV-visible spectrophotometry.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. GB/T 191, Packaging and storage marks GB 4793.1, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1.General requirements GB 4793.9, Safety requirements for electrical equipment for measurement, control and laboratory use - Part 9.Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes GB/T 14710, Environmental requirement and test methods for medical electrical equipment

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply. 3.1 absorbance The ratio of the transmitted light intensity to the incident light intensity is the transmittance; the common logarithm of the reciprocal of the transmittance is called the absorbance. 3.3 carry-over Carrying a test sample reaction by the measurement system to another analyte discontinuity that detects the sample reaction, thereby erroneously affecting the amount of performance of the other test sample. 3.4 stray light Light other than the wavelength that deviates from the normal light path and reaches the detector.

4 Classification

4.1 Instrument type Discrete, mobile. 4.3 Light path form Front splitting or post splitting. 4.4 Colorimetric container type Recycling or single use.

5 Requirements

5.1 Normal working environment 5.1.1 Power supply voltage. 220V ± 22V, 50Hz ± 1Hz. 5.2 Stray light Absorbance is not less than 2.3. 5.3 Absorbance linear range The maximum absorbance of the relative bias within ±5% shall be no less than 2.0. 5.5 Absorbance stability The change in absorbance shall not be greater than 0.01. 5.6 Absorbance repeatability Expressed by coefficient of variation, not exceeding 1.5%. 5.7 Temperature accuracy and volatility The temperature value is within ±0.3°C of the set value, and the fluctuation is not more than ±0.2°C. 5.8 Sample carry-over rate The sample carry-over rate shall not exceed 0.1%. 5.10 Intra-assay precision of clinical projects The coefficient of variation (CV) shall meet the requirements of Table 2. 5.11 Appearance requirements The appearance shall meet the following requirements. 5.12 Environmental test requirements Meet the requirements of the applicable provisions of GB/T 14710. 5.13 Safety requirements Meet the requirements of the applicable provisions of GB 4793.1, GB 4793.9, YY 0648. 5.14 Electromagnetic compatibility requirements Meet the requirements of the applicable provisions of GB/T 18268.1 and GB/T 18268.26.

6 Test methods

6.1 Stray light Using deionized water as a reference, determine a 50 g/L sodium nitrite standard solution at 340 nm (see Annex A for the preparation method); or measure the absorbance of the JB400 cut-off filter at 340 nm using air as a reference, which shall meet the requirements of 5.2. 6.3 Absorbance accuracy Use deionized water as a reference. Determine the absorbance of the potassium dichromate standard solution of which the absorbance is about 0.5 (with deionized water as blank, allowable deviation is ±5%) and 1.0 (with deionized water as blank, allowable deviation of ±5%) at 340 nm on the analyzer. 6.4 Absorbance stability Perform the absorbance stability measurement at any wavelength in the 340 nm and 600 nm to 700 nm wavelength range of the analyzer. The 340 nm measurement solution is Orange G standard solution with an absorbance of 0.5 (with deionized water as a blank and an allowable deviation of ±5%). 6.5 Absorbance repeatability Carry out the absorbance repeatability measurement at 340 nm wavelength of the analyzer. The 340 nm wavelength measurement solution is Orange G standard solution with an absorbance of 1.0 (with deionized water as a blank and an allowable deviation of ±5%). 6.6 Temperature accuracy and fluctuation Place the probe of the temperature tester with an accuracy of not less than 0.1°C, or the same precision and calibrated special temperature test tool provided by the analyzer manufacturer, at the position specified by the manufacturer. After the temperature is stable, the temperature shall be measured every other analyzer interval or 30s. The measurement time is the analyzer's nominal maximum reaction time or 10min. 6.10 Appearance Perform visual inspection, which shall comply with the provisions of 5.11. 6.11 Environmental test Test in accordance with the method specified in GB/T 14710, the results shall comply with the provisions of 5.12. 6.12 Safety requirements Test according to the methods specified in GB 4793.1, GB 4793.9, YY 0648, the results shall meet the requirements of 5.13. 6.13 Electromagnetic compatibility Test according to the method specified in GB/T 18268.1 and GB/T 18268.26, the results shall comply with the provisions of 5.14.

7 Marks and instructions for use

Meet the requirements of GB/T 29791.3.

8 Packaging, transport and storage

8.1 Packaging The analyzer packaging shall meet the following requirements. ......
Source: Above contents are excerpted from the full-copy PDF -- translated/reviewed by: www.ChineseStandard.net / Wayne Zheng et al.


      

Tips & Frequently Asked Questions

Question 1: How long will the true-PDF of English version of YY/T 0654-2017 be delivered?

Answer: The full copy PDF of English version of YY/T 0654-2017 can be downloaded in 9 seconds, and it will also be emailed to you in 9 seconds (double mechanisms to ensure the delivery reliably), with PDF-invoice.

Question 2: Can I share the purchased PDF of YY/T 0654-2017_English with my colleagues?

Answer: Yes. The purchased PDF of YY/T 0654-2017_English will be deemed to be sold to your employer/organization who actually paid for it, including your colleagues and your employer's intranet.

Question 3: Does the price include tax/VAT?

Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countries

Question 4: Do you accept my currency other than USD?

Answer: Yes. www.ChineseStandard.us -- YY/T 0654-2017 -- Click this link and select your country/currency to pay, the exact amount in your currency will be printed on the invoice. Full PDF will also be downloaded/emailed in 9 seconds.

Question 5: Should I purchase the latest version YY/T 0654-2017?

Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version YY/T 0654-2017 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.

How to buy and download a true PDF of English version of YY/T 0654-2017?

A step-by-step guide to download PDF of YY/T 0654-2017_EnglishStep 1: Visit website https://www.ChineseStandard.net (Pay in USD), or https://www.ChineseStandard.us (Pay in any currencies such as Euro, KRW, JPY, AUD).
Step 2: Search keyword "YY/T 0654-2017".
Step 3: Click "Add to Cart". If multiple PDFs are required, repeat steps 2 and 3 to add up to 12 PDFs to cart.
Step 4: Select payment option (Via payment agents Stripe or PayPal).
Step 5: Customize Tax Invoice -- Fill up your email etc.
Step 6: Click "Checkout".
Step 7: Make payment by credit card, PayPal, Google Pay etc. After the payment is completed and in 9 seconds, you will receive 2 emails attached with the purchased PDFs and PDF-invoice, respectively.
Step 8: Optional -- Go to download PDF.
Step 9: Optional -- Click Open/Download PDF to download PDFs and invoice.
See screenshots for above steps: Steps 1~3    Steps 4~6    Step 7    Step 8    Step 9