YY/T 1759-2020 PDF in English
YY/T 1759-2020 (YY/T1759-2020, YYT 1759-2020, YYT1759-2020)
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Name of Chinese Standard | Status |
YY/T 1759-2020 | English | 410 |
Add to Cart
|
0-9 seconds. Auto-delivery.
|
Guide for design and evaluation of primary flexible packaging for medical devices
| Valid |
Standards related to (historical): YY/T 1759-2020
PDF Preview
YY/T 1759-2020: PDF in English (YYT 1759-2020) YY/T 1759-2020
YY
PHARMACEUTICAL INDUSTRY STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
ICS 11.040.01
C 30
Guide for Design and Evaluation of Primary Flexible
Packaging for Medical Devices
ISSUED ON: SEPTEMBER 27, 2020
IMPLEMENTED ON: SEPTEMBER 1, 2021
Issued by: National Medical Products Administration
Table of Contents
Foreword ... 3
1 Scope ... 4
2 Normative References ... 4
3 Terms and Definitions ... 8
4 Content, Method and Application of Design and Evaluation ... 9
Appendix A (informative) Test Methods for Packaging Design and Evaluation
for Reference ... 14
Appendix B (informative) Information on Test Methods for Packaging Design
and Evaluation ... 16
Guide for Design and Evaluation of Primary Flexible
Packaging for Medical Devices
1 Scope
This Standard provides a guide for design and evaluation of primary flexible packaging
for medical devices. This Standard does not involve acceptability criteria.
NOTE: the acceptability criteria are jointly determined by the packaging manufacturer and
the medical device manufacturer.
This Standard is applicable to the design and evaluation of primary flexible packaging
of medical devices aseptically and non-aseptically provided.
2 Normative References
The following documents are indispensable to the application of this document. In
terms of references with a specified date, only versions with a specified date are
applicable to this document. In terms of references without a specified date, the latest
version (including all the modifications) is applicable to this document.
GB/T 451.1 Paper and Board - Determination of Size and Deviation
GB/T 451.2 Paper and Board - Determination of Grammage
GB/T 451.3 Paper and Board - Determination of Thickness
GB/T 454 Paper - Determination of Bursting Strength
GB/T 455 Paper and Board - Determination of Tearing Resistance
GB/T 458 Paper and Board - Determination of Air Permeance
GB/T 464 Accelerated Aging (dry heat treatment) of Paper and Board
GB/T 465.1 Paper and Board - Determination of Bursting Strength after Immersion in
Water
GB/T 465.2 Paper and Board - Determination of Tensile Strength after Immersion in
Water
GB/T 1037 Test Method for Water Vapor Transmission of Plastic Film and Sheet - Cup
Method
GB/T 7706 The Relief Prints for Decorating
GB/T 7742 (all parts) Textiles - Bursting Properties of Fabrics
GB/T 8807 Test Method for Specular Gloss of Plastics
GB/T 8809 Pendulum Impact Resistance of Plastic Film
GB/T 10006 Plastics - Film and Sheeting - Determination of the Coefficients of Friction
GB/T 12914 Paper and Board - Determination of Tensile Properties - Constant Rate of
Elongation Method (20 mm/min)
GB/T 15171 Test Method for Leaks in Sealed Flexible Packages
GB/T 16578 (all parts) Plastics - Film and Sheeting - Determination of Tear Resistance
GB/T 16886 (all parts) Biological Evaluation of Medical Devices
GB/T 17593 (all parts) Textiles - Determination of Heavy Metals
GB/T 19633.1 Packaging for Terminally Sterilized Medical Devices - Part 1:
Requirements for Materials, Sterile Barrier Systems and Packaging Systems
GB/T 19789 Packaging Materials - Test Method for Oxygen Gas Permeability
Characteristics of Plastic Film and Sheeting - Coulometric Sensor
GB/T 21529 Determination of Water Vapor Transmission Rate for Plastic Film and
Sheeting - Electrolytic Detection Sensor Method
GB/T 22819 High Permeable Paper - Determination of Air Permeability
GB/T 22895 Paper and Board - Determination of the Static and Kinetic Coefficients of
Friction - Horizontal Plane Method
GB/T 22898 Paper and Board - Determination of Tensile Properties - Constant Rate of
Elongation Method (100 mm/min)
GB/T 22901 Paper and Board - Determination of Air Permeance (medium range) -
General Method
GB/T 22921 Paper and Board - Determination of Water Vapor Transmission Rate of
Sheet Materials - Dynamic Sweep and Static Gas Methods
GB/T 24218.1 Textiles - Test Methods for Nonwovens - Part 1: Determination of Mass
per Unit Area
GB/T 24218.2 Textiles - Test Methods for Nonwovens - Part 2: Determination of
Thickness
YY/T 0698.6 Packaging Materials for Terminal Sterilized Medical Devices - Part 6:
Paper for Manufacture of Sterile Barrier Systems Intended for Sterilization by Low
Temperature Sterilization Processes or Irradiation - Requirements and Test Methods
YY/T 0698.7 Packaging Materials for Terminal Sterilized Medical Devices - Part 7:
Adhesive Coated Paper for the Manufacture of Sealable Packs for Medical Use for
Sterilization by Ethylene Oxide or Irradiation - Requirements and Test Methods
YY/T 1286.1 Platelets Storage Container Performances - Part 1: Determination of Gas
Permeability of the Film - Differential-pressure Method
YY/T 1432 Test Method for Determination of Heat-sealability of Flexible Webs of
Medical Devices as Measured by Seal Strength
YY/T 1433 Test Method of Hot Seal Strength (hot tack) of Flexible Webs of Medical
Devices
ISO 6588 (all parts) Paper, Board and Pulps - Determination of pH of Aqueous Extracts
ISO 7765-1 Plastics Film and Sheeting - Determination of Impact Resistance by the
Free-falling Dart Method - Part 1: Staircase Methods
ISO 9197 Paper, Board and Pulps - Determination of Water-soluble Chlorides
ISO 15105 (all parts) Plastics - Film and Sheeting - Determination of Gas-transmission
Rate
3 Terms and Definitions
The following terms and definitions are applicable to this document.
3.1 Barrier Requirements
Barrier requirements refer to the demand to promote or restrict moisture, gas, light or
combinations thereof to maintain the necessary level of sterility.
3.2 Durability Requirements
Durability requirements refer to the material characteristics related to the ability of the
package to protect the product.
3.3 Integrity and Seal Requirements
Integrity and seal requirements refer to the ability of the package to prevent accidental
dropping of the contents or entry of foreign substances when the package is not
opened for use.
4.2 Method of Design and Evaluation
When a medical device is characterized and the sterilization method is determined,
then, there are many specific requirements for its packaging. This Standard provides
guidelines for the selection of test methods for the evaluation of these requirements,
the packaging research and development stage and the routine control.
NOTE: product characterization includes quality or weight, geometric shape (length, width,
height and shape) and product composition.
4.3 Description and Application of Test Methods
4.3.1 General principles
Table 1 provides the test methods available for various evaluations suitable for the
research and development stage, and routine control. The provided test methods are
not a test list. The appropriate test method needs to be selected in accordance with
the specific packaging. Other test methods beyond Table 1 may also need to be
selected.
NOTE: see Appendix A for information on some of the available test methods. Please refer
to Appendix B for the description and application of the various test methods in
Table 1.
4.3.2 Tests of packaging research and development stage
Tests of the packaging research and development stage are used to generate
quantitative data on the properties of materials and components. Some test methods
take a long time and are not suitable for process control that requires rapid response;
some test methods are expensive and require special test instruments.
4.3.3 Tests of routine control
Tests of routine control need to be fast, low-cost and easy to carry out in a production
environment. The purpose is not to develop design data, but to ensure that the design
specifications are satisfied. These test methods are not necessarily used to directly
measure the critical values; they may also be used to detect changes in materials,
processes or key characteristics of the product.
4.3.4 Special consideration
The multiple test methods provided in this Standard are not necessarily applicable to
a specific medical device packaging. The selection of a specific test method shall be
based on the relevant characteristics of the medical device product and the purpose
of the test (such as: the research and development stage, and routine control). The
users of this Standard need to consider the applicability of various items and test
methods.
...... Source: Above contents are excerpted from the PDF -- translated/reviewed by: www.chinesestandard.net / Wayne Zheng et al.
|