Non-destructive testing -- Measurement and evaluation of the X-ray tube voltage -- Part 2: Constancy check by the thick filter method
| Valid |
Standards related to (historical): GB/T 41105.2-2021
PDF Preview
GB/T 41105.2-2021: PDF in English (GBT 41105.2-2021) GB/T 41105.2-2021
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 19.100
CCS J 04
GB/T 41105.2-2021 / ISO 16526-2:2011
Non-destructive testing - Measurement and evaluation of the
X-ray tube voltage - Part 2: Constancy check by the thick
filter method
性核查
(ISO 16526-2:2011, IDT)
ISSUED ON: DECEMBER 31, 2021
IMPLEMENTED ON: JULY 01, 2022
Issued by: State Administration for Market Regulation;
Standardization Administration of PRC.
Table of Contents
Foreword ... 3
Introduction ... 4
1 Scope ... 5
2 Normative references ... 5
3 Terms and definitions... 5
4 Principles and devices ... 6
5 Measurement ... 8
6 Measurement report ... 9
Appendix A (Informative) Application example ... 11
Non-destructive testing - Measurement and evaluation of the
X-ray tube voltage - Part 2: Constancy check by the thick
filter method
1 Scope
This document specifies the method for checking the constancy of the X-ray system,
mainly to check the constancy of the X-ray tube voltage and tube current, AND to check
and evaluate the X-ray tube voltage and tube current changes, that may be caused by
the aging of the X-ray tube anode target.
The method, which is specified in this document, is to measure the dose rate, after
defining a filter plate of certain thickness, based on the determined X-ray tube, the filter
plate and the measuring device and the distance.
The method, which is specified in this document, is very sensitive to tube voltage
changes, BUT it cannot measure the absolute value of the X-ray tube voltage. Therefore,
a reference value needs to be given and confirmed, for example in an acceptance test of
an X-ray system.
This document is intended for periodic constancy checks of X-ray systems, by testers
using the thick plate filter method, which is a relatively convenient technique.
This document applies to the constancy check of the X-ray system, after replacement
of components, that may affect the voltage of the X-ray tube.
This document applies to all types of X-ray systems, such as potentiostatic, half-wave
rectified and pulsed, high-voltage generators, which have a tube current greater than 1
mA.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
There are no terms and definitions, that need to be defined in this document.
LV - Leverage factor;
D95 - 95% of the test voltage value, in kilovolts (kV);
D100 - 100% of the test voltage value, in kilovolts (kV).
5.2 Constancy check
At a given voltage, the constancy check of the tube shall be repeated periodically.
Constancy checks are performed, using the same collimator, filter plate (thickness and
material), calibrated dosimeter, through the arrangement corresponding to Figure 1. If
the measured dose value or dose rate value is within the range of the required deviation
T (expressed in %), the constancy check of the ray tube voltage is confirmed; the
deviation T is determined, through calculation by formula (2).
Where:
T - Deviation;
LV - Leverage factor;
D100measured - Measured value at 100% of the given test voltage;
D100nomina - Nominal value at 100% of the given test voltage.
In the reference measurement, all basic parameters (distances D1 and D2, filter material
and thickness, dosimeter or dose rate meter model) shall be recorded; their arrangement
shall remain unchanged, during all subsequent measurements.
Appendix A gives examples.
6 Measurement report
The measurement report shall contain at least the following information:
a) The number of this document, that is, GB/T 41105.2-2021;
b) Model and serial number of the X-ray system;
c) The working conditions of the X-ray system, such as tube voltage (kV), tube
current (mA) and selected focus size, preferably nominal tube voltage and
maximum tube current;
d) Collimator type and size;
...... Source: Above contents are excerpted from the PDF -- translated/reviewed by: www.chinesestandard.net / Wayne Zheng et al.
|