GB/T 1962.2-2001 PDF in English
GB/T 1962.2-2001 (GB/T1962.2-2001, GBT 1962.2-2001, GBT1962.2-2001)
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Name of Chinese Standard | Status |
GB/T 1962.2-2001 | English | 70 |
Add to Cart
|
0-9 seconds. Auto-delivery.
|
Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings
| Valid |
Standards related to (historical): GB/T 1962.2-2001
PDF Preview
GB/T 1962.2-2001: PDF in English (GBT 1962.2-2001) GB/T 1962.2-2001
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 11.040.20
C 31
GB/T 1962.2-2001 / idt ISO 594-2:1998
Conical fittings with a 6 % (Luer) taper for syringes,
needles and certain other medical equipment - Part 2:
Lock fittings
注射器, 注射针及其他医疗器械 6 %(鲁尔)圆锥接头
ISSUED ON: SEPTEMBER 18, 2001
IMPLEMENTED ON: FEBRUARY 01, 2002
Issued by: General Administration of Quality Supervision, Inspection and
Quarantine of the PRC
Table of Contents
Foreword ... 3
ISO Foreword ... 4
1 Scope ... 5
2 Reference standards ... 5
3 Dimensions ... 6
4 Requirements ... 10
5 Test methods ... 12
Foreword
This Standard is identical to ISO 594-2:1998 "Conical fittings with 6 % (Luer)
taper for syringes, needles and certain other medical equipment - Part 2: Lock
fittings".
This Standard was proposed by National Medical Products Administration.
This Standard shall be under the jurisdiction of National Technical Committee
95 on Injector for Medical Purpose of Standardization Administration of China.
Drafting organization of this Standard: Shanghai Centre for Medical Equipment
Quality Supervision and Testing, National Medical Products Administration.
Main drafters of this Standard: Fu Guobao, Zhao Jing.
Conical fittings with a 6 % (Luer) taper for syringes,
needles and certain other medical equipment –
Part 2: Lock fittings
1 Scope
This Standard specifies dimensions, requirements, test methods for conical
lock fittings with a 6 % (Luer) taper for use with hypodermic syringes and
needles and with certain other apparatus for medical use, e.g. transfusion
equipment.
The requirements apply to fittings made of rigid and of semi-rigid materials and
include test methods, but exclude provision for more flexible or elastomeric
materials.
Note
1 It is not practicable to define the characteristics of rigid or semi-rigid materials with
precision, but glass and metal may be considered as typical rigid materials. In contrast
many plastics materials may be regarded as semi-rigid.
2 The Luer lock fitting was designed for use at pressures of the order of 300 kPa or lower.
2 Reference standards
The following standards contain provisions which, through reference in this
Standard, constitute provisions of this Standard. At the time of publication of
this Standard, the editions shown are valid. All standards will be revised. Parties
using this Standard shall explore the possibility of using the latest editions of
the following standards.
GB/T 1031-1995 Surface roughness - Parameters and their values
GB/T 1962.1-2001 Conical fittings with a 6% (Luer) taper for syringes,
needles and certain other medical equipment - Part 1: General requirement
(idt ISO 594-1:1986)
GB 15810-2001 Sterile hypodermic syringes for single use (eqv ISO 7886-
1:1993)
...... Source: Above contents are excerpted from the PDF -- translated/reviewed by: www.chinesestandard.net / Wayne Zheng et al.
|