HOME   Cart(0)   Quotation   About-Us Tax PDFs Standard-List Powered by Google www.ChineseStandard.net Database: 189760 (18 Jan 2025)

GB/T 11913-1989 PDF English


Search result: GB/T 11913-1989_English: PDF (GB/T11913-1989)
Standard IDContents [version]USDSTEP2[PDF] delivered inName of Chinese StandardStatus
GB/T 11913-1989English495 Add to Cart 0-9 seconds. Auto-delivery. Water quality--Determination of dissolved oxygen--Electrochemical probe method Obsolete
BUY with any currencies (Euro, JPY, GBP, KRW etc.): GB/T 11913-1989     Related standards: GB/T 11913-1989

PDF Preview: GB/T 11913-1989


GB/T 11913-1989: PDF in English (GBT 11913-1989)

GB 11913-1989 GB NATIONAL STANDARD OF THE PEOPLE’S REPUBLIC OF CHINA Water Quality - Determination of Dissolved Oxygen - Electrochemical Probe Method APPROVED ON: DECEMBER 25, 1989 IMPLEMENTED ON: JULY 01, 1990 Approved by: State Bureau of Environmental Protection Table of Contents 1 Theme Content and Applicable Scope ... 3 2 Principle ... 4 3 Reagents ... 4 4 Instruments ... 4 5 Procedures ... 5 6 Presentation of Results ... 7 7 Test Report ... 8 Appendix A (Addition)... 9 Water Quality - Determination of Dissolved Oxygen - Electrochemical Probe Method This Standard equivalently adopts the international standard ISO 5814-1984 Water Quality – Determination of Dissolved Oxygen – Electrochemical Probe Method. 1 Theme Content and Applicable Scope 1.1 Theme content This Standard specifies a method for measuring dissolved oxygen in water by an electrode that separates the water sample from the electrochemical cell through a gas- permeable membrane. According to the different types of the used probes, the concentration of oxygen (mg/L), or the percentage of oxygen saturation (% dissolved oxygen), or both may be measured. This method may measure dissolved oxygen with a saturation percentage from 0% to 100% in water. However, most instruments can measure supersaturation values higher than 100%. This method may be used not only for determination in the laboratory, but also for on-site determination and continuous monitoring of dissolved oxygen. This method is suitable for the determination of water with high chromaticity and turbidity. It is also suitable for the determination of water containing iron and substances that can interact with iodine. All the above substances shall interfere with the determination by iodometry. Some gases and vapors like chlorine, sulfur dioxide, hydrogen sulfide, amine, ammonia, carbon dioxide, bromine and iodine can diffuse and pass through the membrane. If these substances are present, they shall affect the measured current and cause interference. The presence of other substances in the sample shall cause the membrane blockage, membrane damage or electrode corrosion, and further lead to interfere with the measured current. These substances include solvents, oils, sulfides, carbonates and algae. 1.2 Scope of application This method is applicable to natural water, sewage and salt water. If it is used to measure salt water such as sea water or harbor water, the salt content shall be checked. 4.1.2 Meter, the scale directly displays the concentration of dissolved oxygen, and (or) the saturation percentage of oxygen or the current microampere. 4.2 Thermometer with a scale division of 0.5°C. 4.3 Barometer with a scale division of 10Pa. 5 Procedures When using the measuring instrument, the instruction manual of the manufacture shall be followed. 5.1 Measurement technology and precautions 5.1.1 Do not touch the active surface of the membrane with hands. 5.1.2 After replacing the electrolyte and membrane, or when the membrane is dry, make the membrane wet. Calibration (see 5.2) can only be performed after the reading is stable. The required time depends on the time required for the consumption of dissolved oxygen in the electrolyte. 5.1.3 When the probe is immersed in the sample, it shall be ensured that no air bubbles are trapped on the membrane. 5.1.4 When the sample is in contact with the membrane of the probe, a certain flow rate shall be maintained to prevent the dissolved oxygen in the sample at that part from being exhausted at the moment of contact with the membrane, and generating the false readings. It shall be ensured that the flow rate of the sample does not cause the reading to fluctuate. In this regard, refer to the instructions of the instrument manufacturer. 5.1.5 For dispersed samples, the measuring container shall be able to be sealed and isolate air; and be equipped with a stirrer (such as an electromagnetic stirrer). Fill the container with the sample to overflow; seal it and then measure. Adjust the stirring speed to keep the reading stable after reaching equilibrium, and must not to entrain air. For flowing samples, such as river channels, it is necessary to check whether sufficient flow velocity can be ensured. If it is insufficient, move the probe back and forth in the water sample; or take out a dispersed sample and measure it according to the method described in the previous paragraph. 5.2 Calibration The calibration procedures are described in 5.2.1 to 5.2.3; but the instruction manual of the instrument manufacture must be referred. 5.2.1 Adjustment Adjust the electrical zero point of the instrument. Some instruments have compensation zero point, so there is no need to adjust. 5.2.2 Check zero point When checking the zero point (need to adjust the zero point, when necessary), the probe can be immersed in each liter of distilled water adding 1g of sodium sulfite (3.1) and about 1mg of cobalt salt (II) (3.2). A stable reading shall be obtained within 10min. NOTE: The new instrument only takes 2~3min. 5.2.3 Calibration close to saturation At a certain temperature, aerate into the water to make the oxygen content in the water saturated or close to saturation. Keep at this temperature for 15min and then determine the concentration of dissolved oxygen; for example, by using iodometry. 5.2.4 Adjust the instrument The probe is immersed in the bottle; and the bottle is completely filled with the sample prepared and calibrated according to the above procedures. Allow the probe to stabilize in the stirred solution for 10min (see 5.2.2 NOTE). If necessary, adjust the instrument reading to the known oxygen concentration of the sample. When the instrument can no longer be calibrated, or the response of the instrument becomes unstable or low (see the manufacturer's instruction manual), the electrolyte or (and) membrane shall be replaced. NOTE: ① If the aeration time and air flow rate required for air-saturated samples have been given in the past experience, Table A1 and Table A3 may be referred to replace iodometry determination. ② Many instruments may be calibrated in air. 5.3 Determination Determine the water to be tested in accordance with the manufacturer's instruction manual. After the probe is immersed in the sample, allow the probe to stay for enough time to make the probe and the water temperature to be measured consistent and to stabilize the reading. Due to different instrument models and different requirements for results, the water temperature and atmospheric pressure shall be checked if necessary. 6.3 Corrected dissolved oxygen concentration of brine samples The solubility of oxygen in water decreases with the increase of salt content. In practical applications, when the salt content (expressed as total salt) is below 35g/L, it may be reasonably considered that the above relationship is linear. Table 1 shows the salt content per 1g/L minus the correction value during correction; that is ΔCs. Therefore, when the salt content in water is ng/L, the solubility of oxygen in water is equal to the corresponding solubility in pure water minus nΔCs. 6.4 Dissolved concentration expressed as saturation percentage This is the actual dissolved oxygen concentration expressed in mg/L, which needs to be temperature corrected if necessary; and divided by the theoretical value given in Tables A1 and A3 and obtain a percentage: 7 Test Report The test report includes the following information: a. Refer to this national standard; b. The measurement result and its expression method; c. Water temperature during sampling and testing; d. Atmospheric pressure during sampling and testing; e. Salt content in water; f. The model of the used instrument; g. Special details that may be noticed during the measurement; h. Optional operation details not specified or considered in this national standard. (measured value) (theoretical value) ......
 
Source: Above contents are excerpted from the PDF -- translated/reviewed by: www.chinesestandard.net / Wayne Zheng et al.